![ENDEVCO 133 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/endevco/133/133_instruction-manual_2406718020.webp)
ENDEVCO
www.endevco.com Tel: +1 (866) ENDEVCO [+1 (866) 363-3826]
IM133, Page 20
®
2.4.9 Sample Set Up
The following information is intended for the first time or infrequent user of this amplifier.
A. Using figure 1 as a reference, perform the following initial setups:
1. Using the "Select Channel" button, select channel 1. (The LED illuminated is the channel
that is being monitored by the DPM.)
2. Using the "Select Function" button, select the appropriate sensor type. While the LED is
flashing for the Input Select function, press the "Edit" button directly above or below the word "Edit"
on the front panel. This will cause the DPM to display either "Chrg" or "Volt". Once the appropriate
sensor type is selected, press the "Select Function" button to save that configuration of input and the
"Select Function" LED will automatically advance to the next function. (Assume a PE accelerometer
is used with a sensitivity of 10.04 pC/g.)
3. With the "Current Exc. (mA)" LED flashing, use the button directly above or below the
word "Edit" to select either 0, 4 or 10 mA. If, in step two above, the configuration was set for a "Chrg"
type input then the current level will not affect the charge-type accelerometer. If a voltage or
ISOTRON accelerometer was selected, then it is acceptable to set the current level to 4 mA as a
starting point. (Only one common current level may be set for all three channels.) (Assume a PE
accelerometer is used, so any selection is acceptable.)
4. Press the "Select Function" button to advance to "Sensitivity (pC or mV/EU)". While the
LED is flashing, enter the correct sensitivity value of the sensor being used. (This is accomplished
with the top and bottom "Edit" buttons to increase or decrease the value. The left and right "Edit"
buttons move the digiton the DPM to be changed. The actual value must be within the range of
0.001 to 9999.) Using the same assumption, input 10.04 on the DPM.)
5. Press the "Select Function" button to advance to "Output Scaling (mV/EU)". While the
LED is flashing, enter in an appropriate value, which must be within the following limits: (The scaling
value entered is valid for 0.001 to 9999 along with the following limitation: 0
≤
Output Scaling /
Sensor sensitivity
≤
1000.
To better understand the correct scaling value to enter, the following questions must be answered:
a. What is the desired full scale output voltage?
b. What is the maximum g range to be measured (in g's or EU)?
c. What is the sensor sensitivity?
Metal Shield
Cover
Unit Front
Power Supply
Filter Modules
Channel 1
Channel 2
Channel 3
Figure 7
(Shown cut
away to
expose the
circuit board
and filter
locations.)