Operation
Chapter 2 cont/…
Operation with the H/T Probe cont/…
During operation, the set value will be displayed for a few seconds and
the display will automatically change back to reading the process value of
temperature. The inlet and outlet temperatures as well as relative
humidity will remain displayed during the operations. While in standby, only
the set value of the inlet temperature will be displayed. This is the only option
which can be displayed in section 2 that can actually be set during
manual operation.
Installing a Pulse Flow Device
To attach a Pulse Flow Device: -
1) Remove the four nuts from the four studs, two of which hold the bracket
containing the stainless steel filter. Remove the SS Filter and bracket.
2) Place four wavy washers on the four studs; followed by the four spacers
provided with the PFD. (The spacers are 3.5cm in length). The spacers should
be made finger tight plus an additional ¼ turn.
3)The PFD should be positioned on the spacers with the on/off switch in the
lower left hand corner, facing the back of the instrument and the power
supply/signal line coming from the bottom of the unit. The PFD should be
pushed tightly against the instrument.
4) The bracket and SS Filter should be placed horizontally across the two
spacers. This may take a little bit of force if the bracket has been distorted at
all 4 wavy washers and 4 self locking nuts (M4 size) should be placed on the
screw threads of the four spacers sequentially and tightened.
5) The cable coming from the bottom of the PFD should be connected to the
15 way D socket adjacent to it .The PFD is now installed.