Chattanooga Group ("Company") warrants that the Hydrocollator Models SS, E-2, and E-1 Heating 
Units ("Products") are free of defects in material and workmanship. This warranty shall remain in effect 
for one year from the date of original consumer purchase. If these Products fail to function during the 
one year warranty period due to a defect in material or workmanship, Company or the selling dealer 
will repair or replace this Product without charge within a period of thirty days from the date on 
which the Product is returned to the Company or the dealer.

All repairs to the Product must be performed by a service center authorized by the Company. Any 
modifications or repairs performed by unauthorized centers or groups will void this warranty.

To participate in warranty coverage, this Product's warranty registration card (included with Product) 
must be filled out and returned to the Company by the original owner within ten (10) business days 
of purchase.

This Warranty Does Not Cover:

Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a 
certified Company service agent.
Defects or damage caused by labor furnished by someone other than Company, the selling dealer or 
a certified Company service agent.
Any malfunction or failure in the Product caused by product misuse, including, but not limited to, 
the failure to provide reasonable and necessary maintenance or any use that is inconsistent with the 
Product User's Manual.

COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL 

OR CONSEQUENTIAL DAMAGES.

Some states do not allow the exclusion or limitation of incidental or consequential damages, so the 
above limitation or exclusion may not apply to you.

To obtain service from Company or the selling dealer under this warranty:

 1. A written claim must be made within the warranty period to the Company or the selling dealer. 
Written claims made to the Company should be sent to:

  4717 

Adams 

Road

  P.O. 

Box 

489

  Hixson, 

TN 

37343 

USA

 

 

Telephone: (423) 870-2281 USA

 

 

Telephone: (800) 361-3661 Canada

 

 

Outside USA: +1 (423) 870-7200

  Facsimile: 

(423) 

875-5497

 2. The Product must be returned to the Company or the selling dealer by the owner.

This warranty gives you specific legal rights and you may also have other rights which vary from state 
to state.

The Company does not authorize any person or representative to create for it any other obligation or 
liability in connection with the sale of the Product. Any representation or agreement not contained in 
the warranty shall be void and of no effect.

THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, 

INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

WARRANTY

Hydrocollator® Heating Units

19

Содержание Chattanooga Group Hydrocollator E-1

Страница 1: ...User Manual 120 220 240 Volt Units Hydrocollator HEATING UNITS E 1 SS E 2 ISO 13485 CERTIFIED ...

Страница 2: ...c is a trademark of Chattanooga Group 2006 Encore Medical Corporation or its affiliates Austin Texas USA Any use of editorial pictorial or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical L P isstrictlyprohibited Thispublicationwaswritten illustrated and prepared for print by Chattanooga Group of Encore Medical L P TABLE OF CONTENTS Hy...

Страница 3: ...to moderate injury or damage to equipment WARNINGS Text with a WARNING indicator explains possible safety infractions that will potentially cause serious injury and equipment damage DANGERS Text with a DANGER indicator explains possible safety infractions that are imminently hazardous situations that would result in death or serious injury NOTE Throughout this manual NOTE may be found These notes ...

Страница 4: ...ce HotPacs which are exposed to contamination hazardous or cleaning materials bodily fluids mold etc CAUTION Grounding the Hydrocollator SS E 2 and E 1 Heating Unit Make certain the unit is electrically grounded by plugging into a hospital grade electrical outlet with a ground terminal receptacle U ground outlet 120 VAC only Follow the National Electrical Code The water temperature in the Hydrocol...

Страница 5: ... the switch is on and unit is operating IMPORTANT Some models 220 240 Volt are fitted with a thermal cut out thermostat which acts as a Non self resetting thermal cut out In the unlikely event of a fault condition this device is designed to open the power circuit between 190 F 14 F 88 C 8 C It is not possible to alter its operating range The operation of the unit may be restored by depressing the ...

Страница 6: ...9 Junction Box J Box Cover 2 Drain 10 IEC connector 3 Overtemp Thermostat 11 Power Cord 4 Thermostat Hydraulic 12 Drain Valve Not Shown 5 Bracket Brkt Bulb Themo A B Assembly 13 Reset Overtemp Button 6 Lid 14 Fuses 7 3 inch Caster 15 Pilot Lamp Red 8 Overtemp Light 16 Pilot Lamp Clear NOTE E 1 E 2 models shown models will vary All 220 240 Volt have this button ...

Страница 7: ...se Bib 31 Replacement Tank 32 Outer Jacket 33 Fuse 10A 34 Lid Kit Handle 35 Water Thief 37 Plastic Feet 38 Pilot Lamp Green 21 Rear View Upside Down J Box Open 26 25 24 27 Right Side Up Back View 17 Hot Pac 23 Side Handle 18 Neck Pac Insert 24 Thermostat Hydraulic 19 Drain Copper Tube 25 Thermostat Control Knob 20 Lid Hinge nuts 26 On Off Switch 21 Lid Knob Handle on back 27 Internal Wiring Harness...

Страница 8: ...ters Temperature Range 160 F 165 F 71 C 74 C Thermal Cut out Temp 190 F 14 F 88 C 8 C Accuracy 10 Heating Up Time Time to 160 F 70 C 3 Hours Cool Down Time Time from 160 F 70 C 2 Hours Safety Class Type B Safety Tests UL 544 Dimensions 20 in H x 18 in L x 16 in W 51 cm H x 46 cm L x 41 cm W Weight 21 lbs 9 5 kg Standard Accessories Includes eight Standard size HotPacs ...

Страница 9: ...1 cm H x 38 cm L x 33 cm W Weight 12 lbs 5 5 kg Standard Accessories Includes two Oversize HotPacs three Standard size HotPacs and 1 Neck Contour HotPac E 1 Model Power Supply 110 120V 220 240 V Rated Frequency 50 60Hz Power Consumption 1000 W Tank Capacity 15 44 Liters Temperature Range 160 F 165 F 71 C 74 C Thermal Cut out Temp 190 F 14 F 88 C 8 C Accuracy 10 Heating Up Time Time to 160 F 70 C 2...

Страница 10: ...te The controls are located in the rear near the bottom of the unit The thermostat controls the temperature of the water It is located near the bottom of the unit on the control panel It has been tested and preset at the factory to normal operating temperature If further adjustment is necessary it can be easily accomplished by turning the adjusting screw located on the thermostat Turning the outer...

Страница 11: ...is seeping 2 Too long a period between cleanings 1 Replace packs 2 Drain and clean Packs Too Hot 1 Thermostat set too high 2 Thermostat failure 1 Adjust thermostat 2 Replace thermostat 1 Thermostat set too low 2 Thermostat failure 1 Adjust thermostat 2 Replace thermostat TROUBLESHOOTING Hydrocollator Heating Units Rust 1 Contact with ferrous metal 2 Deposits on inside of tank 3 Too much chlorine i...

Страница 12: ...ear 22148 19 Drain Copper Tube Part A 21288 20 Lid Hinge Rod Nuts 2 required 21737 21 Lid Knob 21025 22 Lid Rod 22231 Thermo Low Water Kit 20548 26 Overtemp Thermostat Control 20548 27 Wiring Internal Harness 21031 29 DrainValve Ball 1 2 in Replacement 21373 34 Handle Lid Kit 22171 35 Water Thief 21882 37 Feet Plastic 75538 26 On Off Switch 23406 Key numbers refer to Nomenclature section pages 5 a...

Страница 13: ...ord 20568 14 Hose Drain 21635 15 Pilot Lamp Red 22149 16 Pilot Lamp Clear 22148 19 Drain Copper Tube Part A 21288 21 E 2 Lid Handle 20512 Thermo Low Water Kit 20548 27 Wiring Internal Harness 21031 29 DrainValve Ball 1 2 in Replacement 21373 35 Water Thief 21882 26 On Off Switch 23406 Key numbers refer to Nomenclature section pages 5 and 6 Used to hold the Hydraulic Thermostat and Overtemp Thermos...

Страница 14: ...r Cord 20568 15 Pilot Lamp Red 22149 16 Pilot Lamp Clear 22148 20 Lid Hinge Rod Nuts 2 required 21737 28 Rack Stainless Steel 21062 21 Handle Lid 21025 22 Hinge Rod Lid 22668 23 Handle Side Kit 21994 Thermo Low Water Kit 20548 27 Wiring Internal Harness 21031 26 On Off Switch 23406 Key numbers refer to Nomenclature section pages 5 and 6 Used to hold the Hydraulic Thermostat and Overtemp Thermostat...

Страница 15: ...1 Thermo Low Water Kit 20548 27 Wiring Internal Harness 61579 29 Valve Boiler Drain 1 2 I D Chrome 21373 Carrier Fuse 72733 Holder Fuse 72734 34 Handle Lid finished 20531 35 Water Thief 21882 On Off Switch 23406 Lamp Green 22146 Cordset Country Specific 29 Rack Wire 22232 Drain Copper Part B 21289 18 Insert For Neck Contour Packs 21053 32 Tank Replacement w Drain 23385 14 Fuse Anti Surge 6 3 Amp 2...

Страница 16: ...k Stainless Steel 21062 21 Handle Lid 21025 22 Hinge Rod Lid 22668 23 Handle Side Kit 2 21994 Thermo Low Water Kit 20548 27 Wiring Internal Harness 61579 Pilot Lamp Green 22146 11 Cordset Country Specific 18 Insert For Neck Contour Packs 21053 Receptacle Power 74723 Carrier Fuse 72733 Holder Fuse 72734 14 Fuse Anti Surge 6 3 Amp 2 71772 Washer Nylon 21027 Key numbers refer to Nomenclature section ...

Страница 17: ...8 21 E 2 Lid Handle 20512 25 Thermo Low Water Kit 20548 27 Wiring Internal Harness 61579 29 DrainValve Ball 1 2 in Replacement 21373 35 Water Thief 2 21882 26 On Off Switch 23406 Lamp Green Lens 22146 15 Lamp Red Lens 22147 Carrier Fuse 2 72733 Holder Fuse 2 72734 14 Fuse Anti Surge 6 3 Amp 2 71772 22 Lid Rod 90796 Insert For Neck Contour Packs 21053 Rack Wire 22259 Key numbers refer to Nomenclatu...

Страница 18: ...h water Water is constantly lost during operation due to evaporation 4 Chlorine in regular tap water may be present in high enough concentrations to damage your unit If you suspect high levels of chlorine in your Hydrocollator Heating Unit we recommend the addition of a dechlorinator 5 Certain additives such as herbal teas essential oils etc will damage the components of the heating unit including...

Страница 19: ...the deterioration of the pack The pack may also be cleaned by scrubbing the pack with soap and water The simplified method of keeping the pack clean is to keep it immersed in water The Hydrocollator HotPac under constant daily use should have a useful life of at least one year When the pack begins to wear out the filler oozes and leaks through the fabric of the cloth and sections of the pack appea...

Страница 20: ...or failure in the Product caused by product misuse including but not limited to the failure to provide reasonable and necessary maintenance or any use that is inconsistent with the Product User s Manual COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation ...

Страница 21: ......

Страница 22: ...22180D 2006 Encore Medical L P 4717 Adams Road P O Box 489 Hixson TN 37343 1 423 870 2281 1 800 361 6661 1 423 870 7200 1 423 870 2046 chattgroup com ISO 13485 CERTIFIED U S A Outside U S A Fax Canada ...