background image

AEM® Disposable Scissors Insert

Instructions For Use/Care

00555-013    Rev. S 2015/04

1 of 2

EN

   

       

ENCISION recommends placing this document in the Instructions for 
Use/Care Section of your AEM Monitor Operator/Service Manual.

Device Description

Disposable Scissors Insert – For Single Use Only

The Scissors Inserts are designed for use with the ES8000 / ES8200 
series AEM® Handle Assemblies. The 1/2” curved scissors will fi t 
through “true 5mm” surgical trocars and operating scopes. The 3/4” 
scissors and hook scissors will fi t through standard 5.5mm trocars.

Principle of Operation

The AEM Handle Assembly is sized to accept an insulating 
Disposable Sheath (ES0150 series) that allows the surgeon to control 
the amount of exposed electrosurgically active metal at the scissors.

The front and rear handles open and close the working tip (blade 
insert) of the instrument.

Scissors Insert

Shielded Barrel Assembly 
(Insert Inside)

Rotation Knob

Cord Connection

Handle

Trigger

NOTE



This product is rated to 9000 Vp-p. Limit electrosurgical generator 
power setting to 80 Watts or lower (60 Watts for the Conmed 
Aspen Excalibur spray mode). Higher settings may result in 
spurious insulation failure alarms and/or insulation breakdown. 
See Encision AEM Monitor Operation/Service Manual for list of 
compatible electrosurgical generators.



See also Instructions for Use/Care for AEM Instruments, 
Disposable Sheath and Disposable Scissors as applicable.



This product is supplied sterile and is not intended for use more 
than one time. No attempt should be made to reprocess this 
device.

Use with Monopolar Electrosurgery

AEM instruments, in conjunction with an AEM Monitor properly 
connected to the electrosurgical generator (ESU), continuously 
monitor and dynamically manage “stray energy” (insulation failure 
and capacitive coupling) in zones 2  & 3, which are likely out of the 
surgeon’s fi eld of view.

AEM shielding does not cover zone 1, which the surgeon should 
keep in view during instrument activation. As in all applications, 
“misapplied” electrosurgical energy remains the responsibility of the 
attending surgeon.

(Note: Zone 1 equals approximately 1.6cm Tip to Shield)
(Note: Zone 3 equals area of Trocar Cannula)

Zone 1

Zone 2

Zone 3

Laparoscopic procedures should be performed only by surgeons 
having adequate training and familiarity with laparoscopic 
techniques and who are also knowledgeable about anatomy and 
pathology as well as the complications, hazards, risks and benefi ts of 
the procedure. 

Indications/Intended Use

These AEM instruments incorporate the use of AEM technology and 
are intended for use in delivering monopolar electrosurgical energy 
during laparoscopic procedures only. 

AEM instruments are intended for use with the AEM Monitoring 
System and electrosurgical generators having compatibility with the 
AEM Monitor.

Scissors Inserts are intended for use on 

soft

 tissue only.

Contraindications

These instruments are not intended for use when laparoscopic 
electrosurgical techniques are contraindicated.

Instructions For Use

Prior to Use

Thoroughly read these instructions and the instructions in the AEM 
Monitor Operator/Service Manual.

The Disposable Scissors Insert is supplied sterile. Inspect the package 
and product for damage prior to use.

AEM System Setup

See laminated Setup Sheet (00701) when using the ES9005 series 
AEM Cord Adapter and (02678) when using the ES9015 Universal 
Adapter.

WARNING



Laparoscopic surgery may result in gas embolism due to 
insuffl

  ation of gas into the abdomen.



Keep electrical connections dry while in use to prevent potential 
conduction of HF current to the user.



Damaged external insulation AND incorrect setup of the AEM 
Monitor may result in a risk of unintended patient burn, shock or 
fi re hazard. Do not use product having damaged insulation.

CAUTION



Good operating room practice suggests that connections of 
accessories to electrosurgical generators be made only while the 
generator is OFF or on Standby. 



Use these instruments only in conditions that assure adequate 
visualization to minimize risk of misapplied electrosurgical 
energy.



Keep ESU power setting as low as possible for the intended 
purpose to minimize unintended burns.



Damaged internal insulation of the cord and/or instrument, 
or loss of shield continuity, may cause ESU return pad alarms 
triggered by the AEM Monitor’s Fault Indicators. For maximum 
patient safety, discontinue use of the instrument if this occurs.



A singular AEM instrument must be the sole conductor of energy 
to tissue. Do not conduct energy by touching an AEM instrument 
to a second instrument contacting tissue. The second device will 
not be protected from capacitive coupling and insulation failure.



Keep electrosurgical instruments away from the patient and 
operative fi eld when not in use. Accidental activation can result in 
unintended injury to the patient.



See electrosurgical generator manual and AEM Monitor 
Operator/Service Manual for precautions concerning the general 
application of electrosurgical equipment.

Отзывы: