*These checks should include:

      = Recommended

Oxford

®

Presence

23

English

*THESE CHECKS SHOULD INCLUDE:

Initially

Before  Use

Service  Intervals

1. SPREADER BAR:

  Check  the  spreader  bar  for  freedom  of  rotation  and                      

    swing. Check for wear on the central pivot. Check for firm attachment to the   

    boom.

ü

ü

ü

2.  BOOM:

 Check the attachment of the boom to the mast. Make sure there is 

only minimal side movement of the boom and the boom is free to rotate on 

the boom bearing.

ü

ü

ü

3.  MAST: 

Check the operation of the mast-locking device. Make sure the mast 

fully engages into the socket. Check the bottom actuator.

ü

ü

ü

4. ACTUATOR:

 The actuator should require no maintenance other than 

    checking for correct operation and listening for unusual noise.

ü

5.  SMART MONITOR: 

Check the function of the emergency stop button. Inspect 

the hand control socket for correct fitting. Check functioning of the hand control. 

Check the redundant controls and confirm they operate as intended.

ü

ü

ü

6. BATTERIES:

 The batteries are housed in the power pack and should not 

    require maintenance other than the regular charging as detailed in the 

    charging instructions. Check that the connections remain clean.

ü

7.  LEG ADJUSTMENT:

 Check the legs operate in both full extensions (inward/

outward).

ü

ü

ü

8. CASTORS: 

Check all castors for firm attachment to the legs. Check for free 

    rotation of the castor and the wheels.

ü

ü

ü

9.  CLEANING: 

Clean with ordinary soap and water and/or any hard surfacem 

disinfectant. Harsh chemical cleaners or abrasives should be avoided as 

these may damage the surface finish of the lift. Avoid wetting any of the 

electrical parts.

ü

ü

10.  LOAD TEST:

 The load test should be carried out in accordance with the 

manufacturer’s test procedures. It is strongly recommended the testing is 

carried out by an authorised service dealer.

ü

11.  BASE AND WHEELS: 

Ensure base is even and level (all four wheels are 

on the floor).

ü

ü

ü

12.  SLINGS: 

Check for wear and fraying.

ü

ü

ü

13.  LUBRICATION:

 Oil pivot joints, including mast and boom connections, pedal 

assembly, spreader bar joint (only if required).

ü

14. HAND SET: 

Ensure plugged fully into controller.

ü

ü

15.  QUICK RELEASE PIN: 

Ensure the pin is securely fitted before lifting, by 

pressing the barbed end.

ü

ü

16.  FIXINGS:

 Check all nuts, bolts, screws and fasteners for excessive wear 

and for tightness. Replace as required.

ü

Содержание Joerns Hoyer Presence

Страница 1: ...Joerns Hoyer Presence Hoyer Presence To avoid injury read user manual prior to use U S E R M A N U A L Product name to go here Product description to go here U S E R M A N U A L enablelifecare com au...

Страница 2: ...y and Commissioning Instructions 5 4 Safety Precautions 9 5 Standard Control Box Handset 11 6 Smart Monitor Handset 12 7 Operating Instructions 16 8 Removal of Spreader Bar Cradle Systems 19 9 Chargin...

Страница 3: ...The Hoyer Presence Patient Lift 3 Oxford Presence 3 English 1 The Oxford Presence Patient Lift Boom Spreader Bar Electric Actuator Mast Extrusion Rear Castor braked Push Handle Battery Control Pack F...

Страница 4: ...that the transfer of a patient is fully risk assessed and conducted safely over a short distance only The Hoyer Presence is suitable for patients in the SITTING SITTING RECUMBENT and RECUMBENT positi...

Страница 5: ...ACK CHARGING LEAD DESK TOP CHARGER STAND WARNING The Oxford Presence is heavy and will need to be lifted with care You may need assistance to lift the Oxford Presence from the carton Assembly Instruct...

Страница 6: ...is indicated by the label on the side of the mast The electric leg operation will not function unless the mast is fully engaged 4 Turn the locking knob until hand tight 5 Line the handle assembly up t...

Страница 7: ...hand control The red emergency stop button located on the rear of the control box is in the OFF out position Oxford Presence 7 English DISASSEMBLY The lift should not be disassembled unless for servic...

Страница 8: ...r Presence 8 Push the up and down buttons on the hand control and confirm the boom rises and lowers Oxford Presence 8 Englis Push the up and down buttons on the hand control and confirm the boom rises...

Страница 9: ...orrect size and capacity NEVER use a sling which is frayed or damaged ALWAYS fit the sling according to the instructions provided user instructions ALWAYS check the safe working load of the lift is su...

Страница 10: ...DO NOT charge batteries in a bathroom or shower room DO NOT place or store batteries under direct sunlight or near a heat source YOUR lift is for patient lifting DO NOT use it or allow it to be used...

Страница 11: ...h 5 Standard Control Box Handset Emergency Stop Button Emergency Raise Lower Redundant Controls LCD Information Screen Handset Connection Socket Actuator Connection Socket Handset Connection Plug Rais...

Страница 12: ...t Emergency Stop Button Emergency Raise Lower Redundant Controls LCD Information Screen Handset Connection Socket Actuator Connection Socket Handset Connection Plug Raise Lower Buttons Handset Actuato...

Страница 13: ...oard LCD screen Number of patient lift cycles Total work done by actuator Number of lift overloads attempted lifts above the safe working load Number of days since last service interval Data collectio...

Страница 14: ...e symbol will be displayed each time the raise or lower keys are depressed on the user handset To reset the service interval to one year depress the leg open and leg close keys simultaneously for five...

Страница 15: ...The display showing full battery means that the battery is fully charged and the lift is ready for use The display showing a half empty battery indicates that it is time to charge the battery The emp...

Страница 16: ...limit of travel in both directions The hand control plugs into a socket at the base of the control box There is a hook on the rear of the hand control which allows it to be parked on the mast or boom...

Страница 17: ...er upward or down Loop strap slings Oxford Quickfit Full Back Quickfit Deluxe Long Seat Access and Silkfit Safety clip slings Oxford Comfort The 6 point system attaches to the spreader bar hooks via l...

Страница 18: ...reverse this procedure SECURI3 CLIP SYSTEM WARNING WARNING WARNING Joerns Healthcare recommends that slings be checked regularly and particularly before use for signs of fraying or damage DO NOT use s...

Страница 19: ...ft is safe to use FOLLOW THESE INSTRUCTIONS EXACTLY Connection 19 8 Removal of Spreader Bar Cradle Systems Presence 18 7 Removal of Spreader Bar Cradle Systems FOLLOW THESE INSTRUCTIONS EXACTLY Your p...

Страница 20: ...posite direction removing it completely from the boom This will release the spreader bar completely Oxford Presence 1 English REMOVAL CAUTION Be sure to support the weight of the spreader bar cradle b...

Страница 21: ...k from the charger Fit the power pack to the lift and make sure the latch holding the pack in place is fully engaged Click in place The charging of Hoyer electric lifts is simple and straightforward b...

Страница 22: ...ONFIRM the lift is not giving a low battery indicator when the hand control is operated If a low battery is indicated DO NOT use and place on charge immediately Check the operation of the emergency st...

Страница 23: ...enance other than the regular charging as detailed in the charging instructions Check that the connections remain clean 7 LEG ADJUSTMENT Check the legs operate in both full extensions inward outward 8...

Страница 24: ...Width 40 1 inches 1020 mm Legs Closed External Width 26 4 inches 670 mm Legs Closed Internal Width 21 6 inches 550 mm Overall Height of Legs 4 7 inches 120 mm Overall Height of Low Legs option 2 4 in...

Страница 25: ...not condensing Atmospheric pressure 700 to 1060 hPa Temperature 10 C to 50 C Relative humidity 20 to 90 30 C not condensing Atmospheric pressure 700 to 1060 hPa Control Box IPX4 Actuator IPX4 Off Boar...

Страница 26: ...t The disposal of the charging and control unit should not be mixed with general household waste The disposal of batteries should not be mixed with general household waste The disposal of electronics...

Страница 27: ...t wear from normal use These parts such as DC batteries and casters are not covered under the two year warranty but are covered for 90 days after date of delivery Joerns Healthcare s obligation under...

Страница 28: ...are com au At Enable we source quality affordable and innovative products globally and have them available for fast delivery But we do more than just provide products We give the knowledge and support...

Отзывы: