UVX-600G-C Operating Instructions
4
Document no. 10010904 Rev 1.4
QUICKSTART GUIDE
1.
The display range is 00 through 99. The decimal points indicate the LED intensity level. The RED
LED above the display indicates that the intensity level exceeds the threshold setting.
2.
Connect cable to power supply observing correct polarity. Reference wiring diagram.
3.
Apply power; sensor will initialize and perform its power up sequence.
4.
The relative intensity will be displayed. By aiming the sensor away from any objects the display will
indicate 00. Aim the sensor at a white piece of paper and the display will indicate an intensity
measurement. Move the paper further away from the sensor to decrease the intensity level.
5.
Various adjustments may be made to increase or decrease sensitivity; refer to the Selections
Calibration Adjustment and LED Intensity Level. Refer to Section: Threshold, to after threshold
setting.
6.
User programmable parameters are discussed in detail in the following sections.
OPERATION
Power up
Upon power up, the sensor initializes by turning on all segments on the display and sequencing
through red, yellow and green on the status LED located above the display.
Intensity display mode
During normal operation the sensor display will indicate the relative intensity of a target within its field
of view. The range of the relative intensity display is 00 through 99. The decimal points on the
display indicate the LED output intensity. No decimal points indicate low, one decimal point
indicates medium and two decimal points indicate high intensity.
Calibration Adjustment
The calibration feature allows the displayed measurement to be adjusted to the desired value by the
user.
Adjusting the measured value:
1.
Place target in sensor’s field-of-view. The sensor will display the current reading.
2.
Press either key (for less than 3 seconds); while the current reading is displayed the value
Is flashed slowly indicating that the sensor is in the adjustment mode.
3.
Press T/+ to increment the reading and P/- to decrement the reading.
4.
Stop pressing either key and sensor returns to normal operating mode in 3 seconds.
Note:
While adjusting the reading, when a limit is reached the display flashes at a faster rate.
The selected gain is stored in memory and is retained when power is removed.
Local Lock
The local lock feature allows the sensor to be locked out, preventing adjustments by unauthorized