1) As in page 3, connect the camera to the PC.
2) It takes 6 minutes to warm up after power is supplied to the thermal imaging camera (normal temperature is detected after 6 minutes of camera
operation)
3) Setting
-
Camera Connection : After Selecting “
ⓐ
USB port #”, click “
ⓑ
Connect
” to display the thermal image on the screen.(It takes about 6 Min to stabilize
ⓚ
)
-
Blackbody Selection: Select “None
ⓓ
”.
- Monitoring:
“: Click “
ⓕ
Monitoring area” and drag with the left mouse button on the screen to set the area
→
Start monitoring by selecting
“
ⓖ
Done”.
4) Temperature Correction : Put a person with normal body-heat at the measuring position in front of the camera. And a
fter entering “
ⓜ
36.5” degrees at
the reference temperature location. Click “
ⓛ
Set” to adjust the temperature so that the human body temperature is “
ⓘ
36.5”. If you select Fahrenheit
unit, adjust it to “
ⓘ
97.7
℉
” degrees.
After that, you can adjust the temperature through “
ⓗ
Calibration”. When you click the “
ⓝ
Timer” button, temperature calibration is set after 5 seconds.
[Note] If the temperature display “
ⓘ
” is displayed as LOW, click “
ⓛ
Set
” to perform the calibration.
5 / 10
[
Installation method
2/2 : None Blackbody]
ⓑ
ⓐ
ⓘ
36.5
ⓕ
ⓚ
ⓕ
ⓖ
ⓓ
ⓗ
ⓜ
ⓛ
ⓝ