In the ordinary way the value of the measured signal is presented immediately on the display. 
If the signal is too high, the range is overflowed and “oooo” is displayed. If the signal is too 
low then “uuuu” is displayed. Please, change the load of the sensor or set-up again the device. 
 
 

                              ▲ 

EMS600 can be set-up through the keyboard  ◄ ● ► 

                                          ▼ 

Functions of the keyboard: 
 
▲  display of max value (see part 4 : Menu / KEYb / doVn) 
▼  display of min value or function HOLD (see part 4 : Menu / KEYb / doVn) 
◄  Menu input 
►   no function 
●    zero setting (max. till 50 % of the range) 
 
 

 
4. Set-up 
 

The device EMS600 can work with 4 various force sensors. Each sensor has its own number 
from  1  to  4  in  the  device  and  each  number  has  own  parameter  set.    After  switching-on  the 
device displays the number of the last used sensor (e.g. 

Sn1

). Immediately after switching-on 

the  number  can  be  changed  and  then  the  device  operates  with  this  new  sensor  (with  new 
parameter  set).  Of  course,  it  is  not  necessary  to  change  the  number  and  then  the  device 
operates with the last set sensor. 

 

To enter the menu press the key ◄, the device displays 

rnG

 (see  Fig. 5). To move between 

items of the menu use the keys ▼ or ▲, to change the item press the key ●. After pressing ● 
change the value of item  with  ▼ or ▲. To save the set-up value press  again  ●, for leaving 
without saving press ◄. The values are saved even after power is turned off. 
 

Menu description

 (there are default values in brackets) 

 

rnG

 (9999) Range of the display. The range can be set from -9999 to 9999, after overload max 

value the device displays 

oooo

. Min value can’t be set, it is automatically negative value of the 

range. For example if the max value is 1000, then the min value is -1000. By overload of  min 
value the 

uuuu

 will be displayed. 

 

dP

  (0.0)      Position  of  decimal  point.  It  is  possible  to  set-up  one,  two,  three  or  no  decimal 

places.   
 

dIV

 (1)  Division. It is possible to set-up division 1, 2, 5 or 10. 

 

Mr

  (5)  Measurement rate. Possible values: 1/s, 5/s or 20/s. 

 

CAL

  Calibration. The device can be calibrated in 2 points.  

Procedure of calibration: Unload the sensor and choose 

CAL

 in the device menu. Then press 

●.  The  device  displays 

CAL1

  and  then 

0.0

.  Now  set-up  the  requested  value  and  then  again 

press ●. The first point is calibrated. After some seconds the 

CAL2

 and then value which was 

in 

rnG

 setting will be displayed. Load the sensor (it is recommended 50% of nominal value at 

least) and change the value on the display. The display value should be the same as the load of 
the  sensor.  Then  press  again  ●.  If  procedure  of  calibration  was  regular,  the 

ok

  will  be 

Содержание EMS 600

Страница 1: ...1 EMSYST spol s r o Súvoz 111 SK 911 01 Trenčín tel fax 00421 32 7432400 e mail mail emsyst sk EMS 600 Service instructions Trenčín June 2017 ...

Страница 2: ...d cells and torque sensors Signal from the sensor is amplified converted to a number calculated and then presented on a display It is also possible to read the value via USB interface to a PC Fig 1 EMS600 Fig 2 EMS600 Connectors and Power switch Connectors and power switch Display Keyboard Batteries USB Sensor OFF ON ...

Страница 3: ...ection Connection to computer Connect the device to computer with standard USB cable There is USB Mini B connector on the device The connector layout is on the figure 4 Pin Názov Popis 1 VCC 5 VDC 2 D Data 3 D Data x NC No connect 4 GND 0 V Fig 4 USB interface 3 Operating Connect a sensor and switch on the device After switching on the number of the last connected sensor appears on the display e g...

Страница 4: ...ess the key After pressing change the value of item with or To save the set up value press again for leaving without saving press The values are saved even after power is turned off Menu description there are default values in brackets rnG 9999 Range of the display The range can be set from 9999 to 9999 after overload max value the device displays oooo Min value can t be set it is automatically ne...

Страница 5: ...lue for 3 seconds 2 Pressing holds the display for 6 seconds 3 Pressing holds the display permanent Subsequent pressing abort the hold function after the lapse of 6 seconds Note The min a max values are not saved in a FLASH memory after switching off they are lost It is possible to clear the both values manually too During display max or min value press the key until the display turn off The new v...

Страница 6: ...cation driver will be installed on the computer automatically Then operating system assigns a free COM port to the device e g COM5 Via this virtual COM port device communicates For communication use the software EMS Center on the producer s home page EMS Center software is used for communication with the device It is available for free on producer s home page ...

Страница 7: ... batteries 7 Technical parameter Power supply 2 x alkaline battery 1 5V size AA Battery current approx 20 mA Operating time min 40 hours Power supply of sensor 3 VDC Resistance of sensor min 300 Ω Sensitivity of sensor 1 2 5 mV V Display accuracy 4 digit 9999 9999 Communication USB 2 0 Working temperature 0 60 C Covering IP40 Housing material ABS Dimensions 154 x 88 x 28 mm Weight 230 g 8 Manufact...

Отзывы: