HFG01 Operation Manual
York EMC Services
Page 12 of 21 Doc. No. Y0321/DOC/038 Rev. 2
5.2 Supplying power to the HFG01
Figure 2: HFG01 power supply
The HFG01 is powered by an external 24 Vdc power supply unit which connects
to the input socket on the front panel
. Pressing the power switch
with the
24 Vdc unit connected will turn the HFG01 on, which will be confirmed by the
green indicator
. Pressing the power switch again will turn the HFG01 off.
5.3 Operating modes
Figure 3: HFG01 mode selection
The operating mode is selected using the Test Mode switch
. The operation of
each test mode is described in Table 1 below. Note that if the Test Mode switch is
changed whilst a test is running, the test is automatically disabled.
Mode
Test Mode Setting Load Profile Description
Flicker
8.3Hz
Produces Pst value of > 1
(above the pass limit defined in EN/IEC61000-3-3)
1 Hz
Produces Pst value of < 1
(below the pass limit defined in EN/IEC61000-3-3).
Harmonics
Fixed
Single repetitive harmonic-rich current waveform
Fluctuating
Repeating current profile alternating between two distinct
waveforms over a 10 second cycle
Table 1: HFG01 load operation test modes