Interferential 960/Multidyne 970
58
Guidance and manufacturers declaration – electromagnetic immunity
The Primo Interferential 960/Multidyne 970 is intended for use in the electromagnetic
environment specified below. The customer or the user of the 960/970 should assure that
it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
±1 kV differential mode
±2 kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T)
For 0,5 cycle
40%
U
T
(60% dip in
U
T)
For 5 cycles
70%
U
T
(30% dip in
U
T)
For 25 cycles
<5%
U
T
(>95% dip in
U
T)
For 5 sec
<5%
U
T
(>95% dip in
U
T)
For 0,5 cycle
40%
U
T
(60% dip in
U
T)
For 5 cycles
70%
U
T
(30% dip in
U
T)
For 25 cycles
<5%
U
T
(>95% dip in
U
T)
For 5 sec
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the 960/970
requires continued
operation during power
mains interruptions, it is
recommended that the
960/970 be powered
from an uninterruptible
power supply or a
battery.
Power
frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic field should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.