
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE
AND MOBILE RF COMMUNICATIONS EQUIPMENT AND
EMBRACEPLUS
EmbracePlus is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The user of the
EmbracePlus can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communication equipment (transmitter) and the EmbracePlus as
recommended below, according to the maximum output power of the
communication equipment.
Frequency Range and Level: RF wireless communication equipment
Test Frequency
(MHz)
Modulation
Minimum
IMMUNITY
Level (V/m)
IMMUNITY
Level
Applied (V/m)
385
Pulse
Modulation: 18
Hz
27
27
450
Pulse
Modulation: 18
Hz
28
28
710
745
780
Pulse
Modulation:
217 Hz
9
9
129