6.
If the transmitter is simulating an ouput current when the fault occurs, the
transmitter continues to generate the simulated current. To alert you that a fault
exists, the word FAULT appears in the main display, and the display flashes.
8.2.8
Troubleshooting contacting and toroidal conductivity
analyzers when a fault or warning message is showing
Message
Explanation
See section
OuEr rAnGe
Over range, measurement exceeds display lim-
it
SEnSor FAIL
Bad sensor, sensor current is a large negative
number
CAL Error
Calibration error, sensitivity (nA/ppm) is too
high or too low
nEEd 0 CAL
Sensor needs re-zeroing, reading is too nega-
tive
rtd FAIL
Bad temperature reading
rtd FAIL, TEMP HI, and TEMP LO
TEMP HI
Temperature reading exceeds 275 °C (527 °F)
when auto temp is selected
rtd FAIL, TEMP HI, and TEMP LO
TEMP LO
Temperature reading is less than -25 °C (-13 °F)
when auto temp is selected
rtd FAIL, TEMP HI, and TEMP LO
SenSE OPEn
Sense line is not connected
OFFSEt Err
Zero offset during standardization exceeds
programmed limit
FACt FAIL
Unit has not been factory calibrated
CPU FAIL
Internal CPU tests have failed
ROM FAIL
Internal memory has failed
AdC Error
Analog to digital conversion failed
OuEr rAnGE and AMP FAIL
These error messages appear if the sensor current is too high. Normally, excessive sensor
current implies that the sensor is miswired or the sensor has failed.
1.
Verify that the wiring is correct and connections are tight. Be sure to check
connections at the junction box if one is being used.
2.
Replace the sensor membrane and electrolyte solution and clean the cathode if
necessary. See the sensor instruction sheet for details.
3.
Replace the sensor.
SEnSor FAIL
Bad sensor means that the sensor current is a large negative number.
Diagnostics and troubleshooting
88
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...