Figure 6-7: Power supply board wiring
A. From DR board for Traditional Architecture configurations
B. Power Supply Board
4. Disconnect the 14-pin ribbon cable from the power supply board.
5. Remove the two long screws that secure the bracket (9,
) to the
Rosemount Xi enclosure.
6. Hold the AC input wiring to the right and slide the power supply board out of the
Rosemount Xi enclosure.
7. Install the new mounting bracket (9,
) on the new power supply board
).
A new bracket and mounting screws are provided in the replacement kit.
8. Slide the power supply board into the mating slots in the Rosemount Xi enclosure.
Make sure the board is correctly aligned in the slots.
9. Install and tighten the bracket mounting screws.
Two new screws are provided in the replacement kit and should be used if the
existing screws are damaged.
10. Connect the ribbon cable to the power supply board.
) is provided in the replacement kit and should be
used if the existing cable is damaged.
11. Connect the AC power plug to the power supply board.
12. Swing the Rosemount Xi cover up and tighten the four mounting screws.
Reference Manual
Maintenance and service
00809-0100-4889
October 2022
Emerson.com/Rosemount
69