Emerson Process Management GmbH & Co. OHG
6-35
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6
Software Menus
Ch2
Tolerances
DeviatToler.:
Off
Zero Limit:
20.0 %
Span Limit:
20.0 %
6.2.3 Setup Menu
6.2.3.2.2
Setup Calibration Tolerances
This parameter determines whether the tole-
rance check is active during calibration (
On
),
or not (
Off
).
If tolerance check is enabled (
On
), setup the
limits for zero and span gas individually.
Accepted values:
0 ... 100 %
(of the channel´s full range)
Multi-channel unit:
In SELECT COMPONENT, select the chan-
nel to be configured.
Multi-channel unit:
Press
left
to enter SELECT COMPONENT,
to change the settings for another channel.
Setup..
Calibration..
Tolerances..