Quick Start Guide
17
July 2016
Field Communicator or AMS
To view or change the scale factor / k-factor:
1. From the
Home
screen, select
2: Configure
.
2. Select
2: Manual Setup
.
3. Select
3: Profiling
.
4. Select
4: Scale Factor P013
or
4: k-factor P013
, depending on the
Non-Linear Profile selected.
5. Input the factor, and press “
ENTER
” to save it.
6. Press “
SEND
” to update the transmitter.
Rosemount 3490 Series Control Unit
To view or change the scale factor / k-factor:
1. From the
Main Menu
screen, select
SETUP
.
2. Select the transmitter (e.g. “
Tx1: 3107
”).
3. Select
DUTY
, and then select
PV Scale Factor
.
4. Follow the on-screen instructions to edit and save the new setting.
Note
Some flow profiles automatically populate this parameter, and do not allow
editing.
Profile height / Power factor (P014)
Level measurement
This is not used for level measurements. It does not appear on the Field Communicator
unless required for volume or flow measurements.
Volume measurement
When the process value (PV) is a volume measurement from a
standard
non-linear-shaped
tank e.g. an ideal horizontal cylinder or a sphere, use this parameter
to enter the diameter (see
Figure 11 on page 16
).
When the PV is a volume measurement from a
regular-shaped
tank e.g. square or
rectangular, this parameter is not used.
When the PV is a volume measurement from an
irregular-shaped
tank, use this
parameter to enter the maximum height. See also the Special Plot section on
page 19
for defining the
irregular-shaped
tank.
Open channel measurement
When the PV) is a flow rate in a
standard
open channel, this parameter is used as the
power factor (‘pwr’ term) in a flow rate calculation. See
page 14
for selecting a flow
profile.
When the PV is the flow rate in an
irregular-shaped
open channel, use this parameter to
enter the maximum height. See also the Special Plot section on
page 19
for defining
the
irregular-shaped
channel.
Scale Factor P013
1
1
ESC
DEL
ENTER
HELP
w
s
q
a
z
Lock
Shift
@ &
á ü
x
e
d
c
r
f
v
t
g
b
y
h
n
u
j
m
*
-
+
/
.
0
7
4
1
8
5
2
9
6
3
i
k
o
l
FN
(Field Communicator screen)
PV Scale Factr P013
1.000
Esc=Quit =Edit
(Rosemount 3491 screen)
Содержание Rosemount Ultrasonic 3107
Страница 24: ...July 2016 24 Quick Start Guide Figure 13 EC declaration of conformity certificate ...
Страница 25: ...Quick Start Guide 25 July 2016 ...
Страница 26: ...July 2016 26 Quick Start Guide ...