Pump Cover
Figure 10-6:
4.
, push the tubing fitting straight outward
until the fitting slides out of the socket.
Pushing the Tubing Fitting Out
Figure 10-7:
5.
Remove and discard the pump tubing.
6.
Insert the new tubing one end at a time. Tongues on the sides of the gray fittings at
the end of the tube fit into receiving grooves in the pump casing. Push the fitting in
place until it clicks. Gently stretch the tubing over the rollers and insert the other
fitting into the receiving socket on the other side of the pump.
.
Inserting New Tubing
Figure 10-8:
7.
Replace the pump cover.
Maintenance
78
Rosemount TCL