5
Troubleshooting
5.1
Cleaning or replacing the electrodes
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are
indicators that the electrodes may need cleaning or replacement.
1.
Use a mild non-abrasive detergent to clean and remove oil deposits from the
electrode.
2.
Soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution to remove
scale deposits.
Temperature effect on life expectancy: If glass electrode life expectancy is 100% at
25 °C (77 °F), then it will be approximately 25% at 80 °C (176 °F), and approximately
5% at 120 °C (248 °F).
Troubleshooting
Rosemount Hx338+ pH Sensor
13