4
Verify the Rosemount 5300 transmitter
Prerequisites
Open the Rosemount Radar Master and record the serial number and
firmware revision of the device to be tested in the
Verification data
tab of
the workbook. This information can be found in the
Device Properties
window in the
Device
menu in Rosemount Radar Master.
Note
Inches or millimeters should be selected for measurement units (there is a
possible inaccuracy of 1.25 mm (0,05 in.) when inches are used for the
display).
Procedure
1. Configure device for verification.
a) Select
Device
→
Backup Config to File…
and save a backup
file of the current device.
The backup file needs to be reloaded once the verification is
complete.
b) Select
Setup
→
Tank
and configure the following parameters
for the transmitter that will be tested:
•
Probe type: User Defined
•
Probe length: Longest cable 50 cm (20 in)
•
Probe Impedance: 100 Ω
•
Reference Pulse Amplitude: 3000 mV
Quick Start Guide
April 2019
10
Rosemount VeriCase