9.4.6
Sensor Temperature High or Low
The sensor RTD is most likely miswired.
Procedure
1. Check wiring connections.
2. Replace the sensor.
9.4.7
Broken Glass Disabled
The impedance of the pH glass electrode is a strong function of temperature. As
temperature increases, the glass impedance decreases. Because the broken glass fault
message appears when the glass impedance becomes too low, it is important that low
impedance readings be properly connected for the temperature effects. However, there is
a high temperature cutoff beyond which the correction does not work. Once the
temperature exceeds this value, the broken glass fault is automatically disabled.
Important
This warning should never appear in the FCL-02.
9.5
Troubleshooting when no error message is
showing
Problem
See Section
Zero current was accepted, but the current is substantially
greater than 10 nA.
Error or warning message appears while zeroing the sensor
(zero current is too high).
Zero current is unstable.
Sensor can be calibrated, but the sensitivity is significantly
different from 350 nA/ppm.
Sensor can be calibrated, but the current is too low.
Process readings are erratic..
Readings drift.
Sensor does not respond to changes in chlorine level.
Sensor does not respond to changes in chlorine level.
Chlorine reading spikes following rapid change in pH.
Chlorine readings spike following sudden changes in pH
(automatic pH correction).
9.5.1
Zero current is too high.
1. Is the sensor properly wired to the transmitter? See
2. Is the zero solution chlorine free? Take a sample of the solution and test it for free
chlorine level. The concentration should be less than about 0.02 ppm.
Reference Manual
Troubleshoot
00809-0100-3412
May 2019
Rosemount FCL 1056
93
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...