6.4.5
Standardize pH value
You can change the pH value measured by the transmitter to match the reading from a
second or referee instrument. The process of making the two readings agree is called
standardization.
Procedure
1. Place the sensor in the flow cell. Wait until pH readings are stable.
2. Press
MENU
.
The main
Menu
screen appears. The cursor is on Calibrate.
3. Press
ENTER
.
4. Choose the sensor you wish to calibrate.
Sensor 1 is the chlorine sensor. Sensor 2 is the pH sensor.
5. Choose pH.
6. Choose Standardize.
7. Once the reading is stable, measure the pH of the liquid using a referee instrument.
Because the pH of may natural and treated waters depends on temperature,
measure the pH of the sample immediately after taking it. For poorly buffered
samples, determine the pH of a continuously flowing sample from a point as close
as possible to the sensor. Change the reading to match the reading of the referee
instrument.
If the calibration is successful, the screen below is displayed for five seconds. The
display then returns to the screen in step 3. If the calibration is not successful, the
existing calibration data is not changed. A screen appears identifying the error (high
slope, low slope, or offset error). For troubleshooting, see
.
8. To return to the main display, press
MENU
and then
EXIT
.
6.5
Calibration - analog outputs
6.5.1
Trimming analog outputs
Although Emerson calibrates the analog outputs at the factory, you can trim them in the
field to match the reading from a standard milliameter. You can trim both the low (0 or 4
mA) and high (20 mA) outputs
6.5.2
Calibrate analog outputs
Procedure
1. Connect a calibrated milliameter across the output you wish to calibrate. If a load is
already connected to the output, disconnect the load.
Do not put the milliameter in parallel with the load.
2. Press
MENU
.
The main
Menu
screen appears. The cursor is on Calibrate.
Reference Manual
Calibrate
00809-0100-3412
May 2019
Rosemount FCL 1056
73
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...