Figure 3-2: Loose Strap on the Sensor
6. Tighten the strap by hand to gently hold the sensor in place. If
necessary, adjust the position of the sensor to ensure correct
alignment around the circumference of the pipe.
7. If there is an excessive length of spare strap, the excess may be cut
off by using the cutter on the strap tightening tool.
Figure 3-3: Cutting away excess strap
8. Prepare the lanyard kit and decide how it will be positioned.
Wrap the lanyard around the circumference of the pipe. The 7 ft (2
m) length will accommodate a maximum diameter of 24 in (610
mm). When it is not possible to wrap the lanyard around a pipe, find
an alternative attachment point for the lanyard.
9. Thread the bare end of the wire through the loop in the lanyard to
secure it to the pipe.
Quick Start Guide
September 2022
12
Emerson.com/Rosemount
Содержание Rosemount ET310C