Faults menu
The
Faults
menu displays a list of faults affecting the analyzer that have been identified by
the BIST.
If no faults have been identified by the BIST the central area of the
Faults
menu will display
No FAULTS
Figure 11-5: Faults Menu - No Faults Indicated
Procedure
1. Press the
REFR
button to refresh the fault list (B).
2. Use the
Scroll Up
(D) and
Scroll Down
(G) buttons to highlight a fault for further
investigation by the BIT system.
3. Once the fault that you wish to diagnose has been highlighted, press the
SEL
button
to select that fault.
The
Fault Display
screen for the component that is reporting an error will then open.
The
Fault Display
screen will display the parameters of the suspect component.
shows an example of a Fault Display; the precise data that will be
displayed will vary depending upon where the fault has occurred.
Troubleshooting and diagnostics
Reference Manual
October 2019
00809-0100-4512
116
Rosemount CT5100(Ex)
Содержание Rosemount CT5100
Страница 1: ...Reference Manual 00809 0100 4512 Rev AA October 2019 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 16: ...16 ...
Страница 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Страница 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Страница 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Страница 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Страница 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Страница 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Страница 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...
Страница 172: ...ATEX IECEX assessment report summary Reference Manual October 2019 00809 0100 4512 172 Rosemount CT5100 Ex ...