Figure 2-2: Full Rack Installation Requirements
A. Four off
⌀
5.5mm mounting slots
B. Four off M4 mounting holes for customer supplied mounting ears (if required)
C. Three off M4 mounting holes for customer supplied rail (if required)
6. If rigid tubing is being connected to the analyzer further clearance will be required
to allow full access to the connections for installation and maintenance.
7. With rigid tubing fitted the analyzer will not be capable of being slid out of the rack
on the slides.
8. Install the analyzer in a standard 19 in. (482.6 mm) rack system.
Note
Supply of the rack, telescopic slide rails, fitting and securing the analyzer into the
rack is the responsibility of the Customer. (Typical requirement - 2 off telescopic
slides, 24 in. (610 mm) long, 121.3 lb (55 kg) maximum load 2 in. (50.8 mm) over
travel 4 in. (9.6 mm) slide thickness Lock-out Front disconnect.).
9. Installation of telescopic slides on the analyzer will be in accordance with the
manufacturer’s procedure the locations for the slide fixing points are provided in
and
.
Install
Reference Manual
January 2020
00809-0100-4244
24
Rosemount CT4400
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......