4.2 Replacement of Parts
EXPLOSION HAZARD !
Replacement of parts found defective is permitted only by using original
parts, or suitable approved parts, if permitted by the table below!
Violation voids the approvals and may cause explosions!
4.2
Replacement of Parts
Specifi cally for ATEX approved analyzers
Cable glands
NOT FOR analyzers for
vapor recovery applications!
Any ATEX approved
fl
ameproof type without
compound, suitable according EN 60079-14,
e. g.
- 8163/2 A2F M20 by Stahl
Cable glands
for all ATEX approved
variations
Any ATEX approved
fl
ameproof type with
compound, suitable according EN 60079-14,
e. g.
- ADE 1FC ISO 20 No. 05 by CAPRI, or
- 8163/2–PXSS2K M20 by Stahl
Cable entries stopping plugs
Any ATEX approved
fl
ameproof type, e. g.
- 120209 by CAPRI
Male thread: M20 x 1,5
Spare part # ETC01754