Instruction Manual
IB-106-340AC Rev. 1.3
January 2003
Rosemount Analytical Inc. A Division of Emerson Process Management
Operation 3-1
SPS 4000
SECTION 3
OPERATION
3-1 OVERVIEW
This section specifies the requirements to set up
an Oxymitter 4000 calibration and how to verify
the calibration gas flow setup. It also explains
the differences between automatic and semi-
automatic calibrations and how to initiate them.
3-2 CALIBRATION
REQUIREMENTS
a.
Two tanks of precision calibration gas mix-
tures are required. Recommended calibra-
tion gases are nominally 0.4% and 8.0%
oxygen in nitrogen.
Do not use 100% nitrogen as a low gas
(zero gas). It is suggested that gas for
the low (zero) be between 0.4% and
2.0% O2. Do not use gases with hy-
drocarbon concentrations of more
than 40 parts per million. Failure to
use proper gases will result in errone-
ous readings.
In addition to the optional disposable gas
bottles available from Rosemount, two ad-
ditional sources of calibrated gas mixtures
are:
LIQUID CARBONIC GAS CORP.
SPECIALTY GAS LABORATORIES
700 South Alameda Street
Los Angeles, California 90058
213/585-2154
767 Industrial Road
San Carlos, California 94070
415/592-7303
9950 Chemical Road
Pasadena, Texas 77507
713/474-4141
12054 S.W. Doty Avenue
Chicago, Illinois 60628
312/568-8840
603 Bergen Street
Harrison, New Jersey 07029
201/485-1995
255 Brimley Road
Scarborough, Ontario, Canada
416/266-3161
SCOTT ENVIRONMENTAL
TECHNOLOGY, INC.
SCOTT SPECIALTY GASES
2600 Cajon Blvd.
San Bernardino, California 92411
714/887-2571
TWX: 910-390-1159
1290 Combermere Street
Troy, Michigan 48084
314/589-2950
Route 611
Plumsteadville, Pennsylvania 18949
215/766-8861
TWX: 510-665-9344
2616 South Loop West
Suite 100
Houston, Texas 77054
713/669-0469
b.
A check valve is required at the Oxymitter
4000 (between the calibration fitting and the
gas line) to prevent the migration of process
gases down the calibration gas line.
A typical calibration setup for the Oxymitter
4000 is shown in Figure 1-4.