Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
http://www.raihome.com
Section 3
Startup and Calibration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Before Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Edit Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-3
Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-3
Displaying the Factory Set OPLR . . . . . . . . . . . . . . . . . . . page 3-3
Zero/Span Calibration Check . . . . . . . . . . . . . . . . . . . . . . . page 3-4
OVERVIEW
This section describes the necessary adjustments and precautions to be
enacted before the startup of the OPM 4000 system.
BEFORE STARTUP
You must complete the following before startup is attempted.
1. Measure and record flange-to-flange distance for use in future off line
calibration procedures.
2. If you are using a recorder, DAS, etc.,
DO NOT CONNECT THEM
NOW
. Outputs and inputs from other sources should be left off until
system has been completely checked according to the following
instructions in this section. After system operation has been verified,
connect and test external connections.
3. The air purge and weather cover system, transceiver, retro reflector,
and service module must be installed and power applied.
4. Control unit must be installed and wired to the service module and
customers equipment as applicable.
5. All wiring and mechanical installations must be complete. All wiring
must be checked and power applied to both the control unit and the
stack service module.
6. Beam alignment procedure has been completed.
Startup Service
Emerson Process Management is available to assist you; call our Customer
Support Center (CSC) at 1-800-433-6076 for details and to schedule startup.
Read the entire manual
to familiarize yourself with the instrument before attempting startup.
Содержание Rosemount Analytical OPM 4000
Страница 2: ......
Страница 6: ......
Страница 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Страница 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Страница 38: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 16 Figure 2 12 Typical System Wiring ...
Страница 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Страница 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Страница 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Страница 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Страница 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Страница 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Страница 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Страница 110: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 6 Figure D 5 Retrofit System Wiring Diagram ...
Страница 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Страница 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...