Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
2-14
SYSTEM WIRING
INSTALLATION
The following procedures outline the necessary wring connections for the
OPM 4000 system. Terminal identifications for the control unit are shown in
Figure 2-11. Wiring diagrams are shown in Figure 2-12 and the pin
designations are shown in Figure 2-13.
The recommended cable connecting the transceiver to the control room unit is
Belden PN 8778, six individually shielded twisted pair, 22 AWG, PVC
jacketed, 80ºC. Maximum connecting cable length is 1000 feet (305 m) for 22
AWG wire, and 3000 ft (914 m) for 18 AWG wire.
Control Unit
1. Connect terminal 17 on the control unit to terminal 17 on TB 1.
2. Connect terminal 18 on the control unit to terminal 16 on TB 1.
3. Connect shield covering the wires from terminals 17 and 18 of the
control unit to terminal 15 on TB 1.
4. Connect terminal 11 on the control unit to terminal 14 on TB 1.
5. Connect terminal 12 on the control unit to terminal 13 on TB 1.
6. Connect terminal 6 on the control unit to terminal 12 on TB 1.
7. Connect terminal 5 on the control unit to terminal 11 on TB 1.
8. Connect shield covering the wires from terminals 6 and 5 of the control
unit to terminal 10 on TB 1.
9. Connect terminal 9 on the control unit to terminals 9 and 8 on TB 1.
10. Connect terminal 5L on the control unit to terminal 7 on TB 1.
11. Connect terminal 6L on the control unit to terminal 6 on TB 1.
12. Connect shield covering the wires from terminals 5L and 6L of the
control unit to terminal 8 on TB 1.
13. Connect terminal 7L on the control unit to terminal 5 on TB 1.
14. Connect terminal 9 on the control unit to terminal 3 on TB 1.
15. Connect shield covering the wires from terminals 7L and 9 of the control
unit to terminal 2 on TB 1.
Air Flow Switches
1. Connect terminals 5 and 4 on TB 1 to the transceiver air flow switch.
2. Connect shield covering the wires from terminals 5 and 4 on TB 1 to
terminal 2 on TB 1.
3. Connect terminals 1 and 2 on terminal board 2 (TB 2) to the retro
reflector air flow switch.
4. Connect shield covering the wires from terminals 1 and 2 on TB 2 to
terminal 3 on TB 2.
Disconnect and lock out all power before wiring the system.
Содержание Rosemount Analytical OPM 4000
Страница 2: ......
Страница 6: ......
Страница 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Страница 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Страница 38: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 16 Figure 2 12 Typical System Wiring ...
Страница 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Страница 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Страница 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Страница 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Страница 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Страница 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Страница 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Страница 110: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 6 Figure D 5 Retrofit System Wiring Diagram ...
Страница 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Страница 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...