INFRARed ReMoTe CoNTRolleR

1.  Pressing a menu key allows the user access to

calibrate, program, or diagnostic menus.

2.  Press ENTER to store data and settings. Press NEXT to

move from one submenu to the next. Press EXIT to

leave without storing changes.

3

Instruction Sheet

5081C-HT

LIQ_MAN_ABR_5081C-HT/Rev.G

January 2015

FIGURe 1. Infrared remote controller

3.  Use the editing keys to scroll through lists of allowed

settings or to change a numerical setting to the

desired value.

4.  Pressing HOLD puts the transmitter in hold and sends

the output current to a pre-programmed value.

Pressing RESET causes the transmitter to abandon the

present operation and return to the main display.

ReSeT  -

Press  RESET  to  end  the  current

operation and return to the main display.
Changes  will  NOT  be  saved. 

ReSeT  does

NoT return  the  transmitter  to  factory
default settings.

CAl -

Press CAL to access the calibration

menu.

pRoG -

Press PROG to access the program

menu.

dIAG -

Press DIAG to read diagnostic mes-

sages.

Hold    -

Press  HOLD  to  access  the

prompts  used  for  turning  on  or  off  the
hold function.

ARRoW  KeyS -

Use 

é

and 

ê

keys  to

increase or decrease a number or to scroll
through  items  in  a  list.  Use  the 

ç

or 

è

keys to move the cursor across a number.
A  flashing  word  or  numeral  shows  the
position of the cursor.

eNTeR -

Press ENTER to move from a sub-

menu to the first prompt under the sub-
menu.  Pressing  ENTER  also  stores
changes in memory and advances to the
next prompt.

NeXT -

Press NEXT to advance to the next

submenu or to leave a message screen.

eXIT -

Press EXIT to end the current oper-

ation. Changes are NOT saved.

Содержание Rosemount Analytical HART SMART 5081C-HT

Страница 1: ...d maintaining the instrument To ensure their continued operation to the design specifications personnel should read this manual thoroughly before proceeding with installation commissioning operation and maintenance of this instrument If this equipment is used in a manner not specified by the manufacturer the protection provided by it against hazards may be impaired Failure to follow the proper ins...

Страница 2: ...D Dust Ignition Proof Class II III Div 1 Groups E G Type 4X Enclosure Class I Div 2 Groups A D Suitable for Class II Div 2 Groups E G T4 Tamb 70 C explosion proof Class I Div 1 Groups B D Class II Div 1 Groups E G Class III Div 1 Class I Groups B D Class II Groups E G Class III Tamb 65 C max TRANSMITTeR SpeCIFICATIoNS 25 C Measured Range 0 20 000 µS cm Accuracy 0 5 of reading and 0 001 µS cm Repea...

Страница 3: ...RESET to end the current operation and return to the main display Changes will NOT be saved ReSeT does NoT return the transmitter to factory default settings CAl Press CAL to access the calibration menu pRoG Press PROG to access the program menu dIAG Press DIAG to read diagnostic mes sages Hold Press HOLD to access the prompts used for turning on or off the hold function ARRoW KeyS Use é and ê key...

Страница 4: ...OutPut OUtPUt 1 Not available with display in Neutral Salt Cation or Resistivity modes OUtPUt Current output menu header 4MA 4mA current output setpoint 20MA 20mA current output setpoint HoLd Current output on hold FAULt Fault condition current output setting dPn Current output dampening time tESt Current output test value tEMP Temperature menu header tAUtO Automatic temperature compensation tMAn ...

Страница 5: ...smitter measurement and configuration variables The user can read raw data final data and program settings FIGURe 3 AMS configuration screen using HART and can reconfigure the transmitter from anywhere in the plant The figures below show two of the many configura tion and measurement screens available using HART AMS FIGURe 4 AMS measurement screen using HART ...

Страница 6: ...Cust Temperature C or F temp C ___________ Output mA or output Cur ___________ Security Code code 000 ___________ Custom Curve SEtUP CuSt ___________ Reference temperature t ref 25 0 C ___________ Calibrate Menu Cell constant CELL Const 1 00 ___________ Temperature slope tEMP slOpe 2 000 ___________ Output Calibration OUtPUt CAL ___________ diagnose Menu diagnose Each segment displays SAMple ReAdI...

Страница 7: ...actory terminated PN 23747 00 and unterminated PN 9200275 connecting cable are available The use of fac tory terminated cable is strongly recommended To pre pare unterminated cable for use follow the instructions in the sensor instruction manual Recommended Transmitter Settings Assumes 1 0 cell constant Software Setting process Stream Type Contacting Conductivity in the range 20 to LInEAr 20 000 m...

Страница 8: ... the ribbon cable between the display board and the CPU board Rotate the display The black infrared sensor will be at the top of the display INSTAllATIoN See Figure 7 1 Although the analyzer is suitable for outdoor use do not install it in direct sunlight or in areas of extreme temperatures 2 Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are ...

Страница 9: ... 00 F ROSEMOUNT ANALYTICAL Tamb 20 C TO 65 C I S SUPPLY Ui 30 VDC Ii 200 mA Pi 0 9 W Ci 0 µF Li 0 µH SIGNAL INPUT Uo 6 51 V Io 400 mA Po 642 mW Ci 24 1 µF Li 0 µH 1180 MODEL 5081 C HT 73 IECExBAS09 0159X Baseefa03ATEX0099X Ex ia IIC T4 Ga II 1 G FIGURe 8 ATeX label Information ...

Страница 10: ...10 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 9 FMRC explosion proof Installation ...

Страница 11: ...11 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 10 FM Hazardous Area Approval label ...

Страница 12: ...BSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2 13 14 12 13 UNCLASSIFIED AREA D2 12 DIVISION 2 WIRING METHOD PER THE NEC EXCLUDING NONINCENDIVE FIELD WIRING 13 INSTALATION TO BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE D1 OUTPUT PARAMETERS It Vt TABLE II MODEL 5081 C HT TB1 1 THRU 12 Ci µ F Li mH 30 300 1 0 MODEL NO 275 Voc max OUT Vdc Isc max OUT mA ENTIT...

Страница 13: ...13 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 12 CSA Hazardous Area Approval label ...

Страница 14: ...14 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 13 CSA Intrinsically Safe Installation ...

Страница 15: ...ual temperature use the arrow keys then press ENTER 11 If you selected CondUC or rStvty in step 2 you will see a CoMp Temperature Compensation type screen Use the arrow keys to select desired temperature compensation lineAr linear n SAlt neutral salt CatIon cation conductivity or none raw or uncompensated conductivity Press ENTER If you are in lineAR mode you can now enter a particular temperature...

Страница 16: ...marks are the property of their respective owners The contents of this publication are presented for information purposes only and while effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express or implied regarding the products or services described herein or their use or applicability All sales are governed by our terms and conditions which ar...

Отзывы: