14
Configuration is all accomplished through a series of
menus located within the set mode menu. To access
these set mode menus the ACCESS keypad is
pressed
T
WICE
in
R
APID
succession.
Once inside the Set mode menu, use the scroll key-
pad to scroll through the menu list. When the menu
desired is displayed, release the scroll keypad.
To enter the submenus press the SELECT keypad. If
the submenu allows editing, the item will flash that can
be edited. If not, use the scroll keypad to scroll through
the next list of submenus. SELECT will enter this sub-
menu and if it is editable, the field will flash.
To exit the menu and SAVE the new value, press the
ENTER keypad.
To exit the menu without saving the edited value, press
the PV keypad to jump out of the set menu program with
out saving value. To change other parameters will
require re-entering the set menu program.
Figure 3-4 explains the various fields surrounding the
Primary process on the LC display.
Table 3-1 describes the functions accessible with the
eight (8) keypads, the number of times to press the
keypad to access, and its function when used with the
select keypad and set menu.
Table 3-2 and Table 3-3 describe the meaning of the
various mnemonics used on the display. They are cat-
egorized by their use in either menus or as process
information.
MODEL 1054B C
SECTION 3.0
DESCRIPTION OF CONTROLS
SECTION 3.0
DESCRIPTION OF CONTROLS
3.1 KEYBOARD FUNCTIONS.
All operations of the
Model 1054B microprocessor Analyzer are controlled
by the eight (8) keypads on the front of the instrument.
These keypads are used to do the following:
1.
Display parameters other than the primary param-
eter.
2.
Edit setpoints for alarms, set up specific output
current value for simulation, calibrate, tempera-
ture, conductivity, etc.
3.
Configure display for temperature units, for auto-
matic temperature compensation, alarm usage,
setting timer functions, security, and output range.
To view and not change parameters other than the pri-
mary parameter requires only a simple keystroke rou-
tine. As shown in Figure 3-1, a single keypress
accesses the lower function printed on the keypad.
Quick, double keypresses access the top function
printed on the keypad.
Editing any of these parameters requires one more
operation. After displaying the value associated with
the parameter selected, press the SELECT keypad.
As seen in Figure 3-2, this will display the numerical
value, and the first digit will be flashing to indicate this
value may be edited.
All changes to the operating program that set-up the
instrument display are made through the set menu pro-
gram. See Figure 3-5 at the end of this section.
FIGURE 3-1.
Function Select on Keypad.
Single press of the keypad will access the
present Conductivity reading. Read only.
1. Press Key.
2.
AdJ
shows briefly.
3. Numbers show with digit flashing.
1. Press Key twice.
2.
Lo
shows briefly.
3. Zero point is displayed.
SELECT
ZERO
ALARM
1
OUTPUT
PV
FIGURE 3-2
. Accessing Editing Function.
1. Press twice in rapid succession.
2. See
SEt
on display. Confirms entry
into set mode menu.
3. First menu item is displayed.
Analyzer now ready to configure.
4. Use the SCROLL keypad to rotate
through the available menus.
ACCESS
ENTER
SEt
Cin
ç
FIGURE 3-3
. Accessing Configuration Menus
Quick double press will access the current
output current value in mA or %. Read only.