119
Reference Manual
00809-0300-4750, Rev CA
Advanced Diagnostics Configuration
March 2016
Advanced Diagnostics Configuration
6.5.1
Turning ground/wiring fault on/off
The
ground/wiring fault detection
diagnostic can be turned on or off as required by the
application. If the advanced diagnostics suite 1 (DA1 Option) was ordered, then the
ground/wiring fault detection
diagnostic will be turned on. If DA1 was not ordered or
licensed, this diagnostic is not available.
6.5.2
Ground/wiring fault parameters
The
ground/wiring fault detection
diagnostic has one read-only parameter. It does not have
any configurable parameters.
Line noise
The
line noise
parameter shows the amplitude of the line noise. This is a read-only value. This
number is a measure of the signal strength at 50/60 Hz. If the
line noise
value exceeds 5 mV,
then the
ground/wiring fault
diagnostic alert will activate.
6.6
High process noise detection
The
high process noise
diagnostic detects if there is a process condition causing an unstable
or noisy reading that is not an actual flow variation. A common cause of high process noise
is slurry flow, like pulp stock or mining slurries. Other conditions that cause this diagnostic
to activate are high levels of chemical reaction or entrained gas in the liquid. If unusual noise
or flow variation is seen, this diagnostic will activate and deliver an alert. If this situation
exists and is left without remedy, it will add additional uncertainty and noise to the flow
reading.
LOI menu path
Field Mount: Diagnostics, Diag Controls, Ground/Wiring
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 2, 1, 1, --
(1)
1.
This item is in a list format without numeric labels.
Device dashboard
2, 2, 5, 1, 3
LOI menu path
Field Mount: Diagnostics, Variables, Line Noise
Wall Mount: XMTR INFO
Traditional Fast Keys
1, 2, 4, 3
Device dashboard
2, 2, 5, 4, 1
Содержание Rosemount 8750W
Страница 2: ......
Страница 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Страница 158: ...146 Reference Manual 00809 0300 4750 Rev CA Digital Signal Processing March 2016 Digital Signal Processing ...
Страница 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Страница 208: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 196 Product Certifications ...
Страница 209: ...197 Product Certifications Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA ...
Страница 212: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 200 Product Certifications ...
Страница 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Страница 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Страница 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Страница 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Страница 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Страница 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Страница 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Страница 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Страница 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Страница 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Страница 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Страница 263: ......