A. OUT_D = Discrete output that signals a selected alarm condition
B. OUT = The block output value and status
The Analog Input (AI) function block processes field device measurements and makes
them available to other function blocks. The output value from the AI block is in
engineering units and contains a status indicating the quality of the measurement. The
measuring device may have several measurements or derived values available in different
channels. Use the channel number to define the variable that the AI block processes.
The AI block supports alarming, signal scaling, signal filtering, signal status calculation,
mode control, and simulation. In Automatic mode, the block’s output parameter (OUT)
reflects the process variable (PV) value and status. In Manual mode, OUT may be set
manually. The Manual mode is reflected on the output status. A discrete output (OUT_D)
is provided to indicate whether a selected alarm condition is active. Alarm detection is
based on the OUT value and user specified alarm limits.
lists the AI block parameters and their units of measure, descriptions, and
index numbers.
Reference Manual
Analog-Input Block
00809-0100-4530
April 2019
Reference Manual
457
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...