Install
Install the sensor in a flowing sample. Keep the sample flow as constant as
possible at a value within the following limits:
Sample flow unit
Flow limits
Flow through
1 to 5 gpm (3.8 to 19 L/min)
Open channel
1 ft/sec (0.3 m/sec)
Low flow cell (PN 24091-00)
8 to 15 gph (20 to 47 L/hr)
Low flow cell (PN 24091-01)
2 to 3 gph (120 to 190 mL/min)
Figure 2-1: Sensor Orientation
Install sensor within 45 degrees of vertical.
Quick Start Guide
September 2019
8