4.1
Sensor Removal
Remove the sensor from the ball valve as follows:
1.
Be certain system pressure at the sensor is below 64 psig (542 kPa) before proceeding
with the retraction. It is also recommended that the personnel wear a faceshield and have
a stable footing.
2.
Push in on the sensor using the top of the J-box and slowly loosen the hex nut (B) of the
process end male connector (A). (Refer to Figure 4-2).
3.
When the hex nut is loose enough, slowly ease the sensor back completely until the
retraction stop collar is reached.
4.
Close the ball valve slowly. If there is resistance it may be hitting the sensor. Double check
that the sensor has been retracted to the retraction stop collar.
5.
The Male Connector Body (A) may now be completely unthreaded from the reducing
coupling and the sensor removed for servicing.
Note:
With the ball valve fully closed, some residual process fluid may leak from the connectors
MNPT pipe threads. This leakage is normal and to be expected.
4.2
Sensor Maintenance
4.2.1
Electrode Cleaning
If the electrode is coated or dirty, clean as follows:
1.
Remove the sensor from process as previously instructed in Section 4.1.
2.
Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the
dirt or coating, go to Step 3. (Detergents clean oil and grease; and acids remove scale.)
3.
Wash the glass bulb in a strong detergent solution and rinse it in clean water. If this does
not clean the glass bulb, go to Step 4.
Section 4: Maintenance
WARNING
System pressure may cause the sensor to blow out with great force unless care is taken during removal.
Make sure the following steps are adhered to.
Failure to withdraw the sensor completely may result in damage to the sensor when the valve is closed.
CAUTION
WARNING
Before removing the sensor from the ball valve, be absolutely certain that the ball valve is fully closed.
Leakage from the male connector threads may indicate that the male connector is still under pressure.
Leakage through a partially open valve could be hazardous.
Maintenance
13
Reference Manual
Maintenance
00809-0100-3385
February 2019
Содержание Rosemount 385
Страница 1: ...Rosemount 385 pH ORP Sensors Reference Manual 00809 0100 3385 Rev AA February 2019 ...
Страница 2: ...hasgkas ...
Страница 4: ......
Страница 6: ...Table of Contents Reference Manual February 2019 00809 0100 3385 ii Table of Contents ...
Страница 13: ...Installation 7 Reference Manual Installation 00809 0100 3385 February 2019 Figure 2 2 Dimensional Drawing ...
Страница 16: ...10 Installation Installation Reference Manual February 2019 00809 0100 3385 ...
Страница 18: ...Startup and Calibration Reference Manual February 2019 00809 0100 3385 12 Startup and Calibration ...
Страница 26: ...20 Troubleshooting Troubleshooting Reference Manual February 2019 00809 0100 3385 ...
Страница 28: ...22 Accessories Accessories Reference Manual February 2019 00809 0100 3385 ...
Страница 30: ...24 EC Declaration of Conformity EC Declaration of Conformity Reference Manual February 2019 00809 0100 3385 ...
Страница 31: ...Reference Manual EC Declaration of Conformity 00809 0100 3385 February 2019 EC Declaration of Conformity 25 ...
Страница 32: ...26 China RoHS Table China RoHS Table Reference Manual February 2019 00809 0100 3385 ...