Installation
5
Section 2: Installation
Reference Manual
Installation
00809-0100-3386
February 2019
2.1
Unpacking and Inspection
Inspect the outside of the carton for any damage. If damage is detected, contact the carrier
immediately. Inspect the instrument and hardware. Make sure all items in the packing list are
present and in good condition. Notify the factory if any part is missing.
2.2
Mechanical Installation
The Rosemount 385+ is available in two versions, the retraction version (Code 02) and the
insertion/submersion version (Code 03, & 04). Please refer to the appropriate section below.
2.2.1
Retraction Option (Code option 02)
The Rosemount 385+ sensor may be installed through a weldalet or in a pipe tee or “Y”, as shown
in Figure 2-1, when used with a ball valve. Insert the end of the sensor to a depth sufficient to ensure
that the glass bulb is continuously wetted by the process fluid.
The sensor can also be inserted directly into the process without the use of a ball valve for
applications not requiring continuous operation during sensor maintenance (Figure 2-2).
Allow sufficient room for safe retraction and insertion of the sensor. personnel should have room
for stable footing while performing removal or insertion of the sensor.
The sensor must be mounted within 10-90 degrees of the horizontal with the tip pointed
downward, thus keeping air bubbles out of the pH sensitive glass bulb. Bubbles settled in the glass
bulb disrupt the electrical continuity between the pH sensitive glass and the silver/silver chloride
measuring element.
If the retraction version is to be installed without a ball valve follow the installation procedure for
insertion service (Section 2.2.2). Perform the following steps for sensor installation through a ball
valve:
1.
Carefully remove the liquid filled rubber boot which protects the glass electrode and keeps
the liquid junction wet during shipping and storage. Discard the liquid and boot. Make
sure the lubricated O-ring is in place in the groove inside the male connector on the sensor
body (Figure 4-1, item A).
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
Save the original packing cartons and materials as most carriers require proof of damage due to mishandling,
etc. Also, if it is necessary to return the instrument to the factory, you must pack the instrument in the same
manner as it was received. Refer to Section 6.0 for instructions.
NOTICE
Buffer solution, in the vinyl boot, may cause skin or eye irritation.
CAUTION
Содержание Rosemount 385+ -02
Страница 1: ...Rosemount 385 pH ORP Sensors Reference Manual 00809 0100 3386 Rev AA February 2019 ...
Страница 2: ...hasgkas ...
Страница 4: ......
Страница 6: ...Table of Contents Reference Manual February 2019 00809 0100 3386 ii Table of Contents ...
Страница 30: ...24 Maintenance Maintenance Reference Manual February 2019 00809 0100 3386 ...
Страница 36: ...30 Accessories Accessories Reference Manual February 2019 00809 0100 3386 ...
Страница 38: ...32 EU Declaration of Conformity EU Declaration of Conformity Reference Manual February 2019 00809 0100 3386 ...
Страница 39: ...Reference Manual EU Declaration of Conformity 00809 0100 3386 February 2019 EU Declaration of Conformity 33 ...
Страница 40: ...34 China RoHS Table China RoHS Table Reference Manual February 2019 00809 0100 3386 ...