Adjust the probe length
Procedure
1. Determine L, the desired probe length.
L, desired probe length:
/
2. Determine n, the number of middle segments needed for the desired probe length.
See
n, number of middle segments:
Q
Q
3. Calculate Y, the length of the bottom segment. See
.
Y, length of bottom segment:
<
4. Continue as follows:
Length of bottom segment (Y)
Action
Y < 0.4 in. (10 mm)
<
•
Continue with step 7. Do not use the
bottom segment.
Y ≥ 0.4 in. (10 mm)
<
•
Continue with step 5 and cut the bottom
segment.
Y = 31.5 in. (800 mm)
<
a. Add one extra middle segment to the
calculated n.
b. Continue with step 7.