SECTION 2.0

MECHANICAL INSTALLATION

2.4 INSTALLATION INSPECTION

Prior to operating the sensor and attempting to stan-
dardize the instrument, inspect the sensor installation
as follows:

1. Verify that the sensor is securely attached to the

wall, pipe, or other mounting surface.

2. Check all plumbing connections for leaks. If any are

evident, repair them.

3. Check all electrical connections for security. Make

sure all wires are properly dressed so they do not
short or ground. Make sure the instrument is wired
as illustrated in Figure 2-3 through 2-10.

4. If all of the above have been satisfied, proceed to

the next section.

FIGURE 2-3. Model 320HP-50/51/54 Wiring to MODELS 1054/A/B, 2054, 2081, 1181, 1003, 1050

DWG. NO.

REV.

40320HP03

A

Содержание ROSEMOUNT 320HP

Страница 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Страница 2: ...High Purity pH Sensor Instruction Manual PN 51 320HP rev J May 2006 Model 320HP ...

Страница 3: ...rstand any of the instructions contact your Rosemount repre sentative for clarification Follow all warnings cautions and instructions marked on and supplied with the product Inform and educate your personnel in the proper installation operation and maintenance of the product Install your equipment as specified in the Installation Instructions of the appro priate Instruction Manual and per applicab...

Страница 4: ...mation 2 2 0 MECHANICAL INSTALLATION 3 2 1 General 3 2 2 Unpacking 3 2 3 Preparation for Installation 3 2 4 Installation Inspection 5 3 0 OPERATION 11 3 1 General 11 3 2 Calibration 11 4 0 MAINTENANCE AND TROUBLESHOOTING 12 4 1 General 12 4 2 Preventative Maintenance 12 4 3 Troubleshooting 13 4 4 Maintenance 14 5 0 PARTS LIST 19 5 1 General 19 6 0 RETURN OF MATERIALS 21 6 1 General 21 6 2 Warranty...

Страница 5: ...l 320HP 58 Wiring to Models 54e 81 3081 4081 and 5081 8 2 8 Model 320HP 55 Wiring to Model 1055 22 32 9 2 9 Model 320HP 58 Wiring to Model 1055 22 32 9 2 10 Model 320HP Wiring to Model 1055 01 10 22 10 3 1 Buffer Cup Fixture 11 4 1 Preamplifier Check 14 4 2 Reference Electrode Check 16 4 3 Glass Electrode Test Method 17 4 4 Temperature Compensator Check 17 5 1 Model 320 HP Front View and Side View...

Страница 6: ...ust be provided for this static charge otherwise it can be easily picked up by the high impedance pH electrode causing excessive noise in the measuring signal The well grounded metallic flow cell in the Model 320HP High Purity pH Sensor surrounds the pH electrode and provides an effective discharge path to ground for any static charges It also serves as an electronic shield against outside interfe...

Страница 7: ...d Pt100 T C for use with Models 1054A B 2054 2081 requires cable PN 9200000 55 No Preamp and Pt100 T C for use with Models 54 54e 1055 81 3081 4081 5081 Xmt and SCL P requires cable PN 661 646983 56 Preamp and Pt100 T C for use with SCL P requires cable PN 661 892884 57 Preamp and Pt100 T C for use with Model 2700 requires cable PN 661 898695 58 Preamp and Pt100 T C for use with Models 54 54e 81 3...

Страница 8: ...eads are not present on the junction surface the liquid junction should be replaced PN 2001317 2 Lubricate the o ring on the reference cell using the o ring lubricant supplied PN 2001928 Place the ref erence cell back on the flow cell Rotate it counter clockwise until the ears seat in the retainer brackets 3 Install sensor panel so the process flow will enter at the bottom of the cell and exit fro...

Страница 9: ...DEL 320HP SECTION 2 0 MECHANICAL INSTALLATION FIGURE 2 2 Model 320HP Dimensional Drawing DWG NO REV 4320HP01 C MILLIMETER INCH 1 Shipped loose customer install and wire For electrical wiring see Table above ...

Страница 10: ...her mounting surface 2 Check all plumbing connections for leaks If any are evident repair them 3 Check all electrical connections for security Make sure all wires are properly dressed so they do not short or ground Make sure the instrument is wired as illustrated in Figure 2 3 through 2 10 4 If all of the above have been satisfied proceed to the next section FIGURE 2 3 Model 320HP 50 51 54 Wiring ...

Страница 11: ...6 MODEL 320HP SECTION 2 0 MECHANICAL INSTALLATION FIGURE 2 4 Model 320HP 55 Wiring to Models 54 54e 81 3081 4081 and 5081 DWG NO REV 4320HP04 C ...

Страница 12: ...MODEL 320HP SECTION 2 0 MECHANICAL INSTALLATION 7 FIGURE 2 6 Model 320HP 58 Wiring to Xmt P HT 10 FIGURE 2 5 Model 320HP 55 Wiring to Xmt P HT 10 ...

Страница 13: ...8 MODEL 320HP SECTION 2 0 MECHANICAL INSTALLATION FIGURE 2 7 Model 320HP 58 Wiring to Models 54e 81 3081 4081 and 5081 DWG NO REV 4320HP09 A ...

Страница 14: ...MODEL 320HP SECTION 2 0 MECHANICAL INSTALLATION FIGURE 2 9 Model 320HP 58 Wiring to Model 1055 22 32 FIGURE 2 8 Model 320HP 55 Wiring to Model 1055 22 32 9 ...

Страница 15: ...MODEL 320HP SECTION 2 0 MECHANICAL INSTALLATION FIGURE 2 10 Model 320HP Wiring to Model 1055 Pipe Wall Mount 10 MODEL 320HP 58 MODEL 320HP 55 ...

Страница 16: ...fer from the cup fixture and rinse with clean potable water 7 Fill buffer cup half full with 4 pH buffer and allow 5 min utes for stabilization of the pH reading Adjust the analyz er s span of operation for a 4 pH reading 8 Remove the 4 pH buffer and rinse thoroughly with potable water Fill the buffer cup fixture half full with 10 pH buffer and allow 5 minutes for stabilization Verify the pH read ...

Страница 17: ...2 If the reference needs recharging turn the reference cell to unlock it from the bracket then pull straight out Remove the junction 3 Empty the electrolyte from the reference cell and flush it with deionized water 4 PIace the reference cell on a flat surface window side down Add deionized water to half fill the cell Slowly add new crystals PN 9210243 until approximately 90 full Top off with deion...

Страница 18: ... Correct wiring shorted mis wired Model 1002 1003 only Broken glass electrode Replace glass electrode Plugged liquid junction Replace liquid junction Drain and source leads between Correct wiring preamp and analyzer open broken wires Model 1002 1003 only Broken reference electrode Replace electrode and recharge ref cell Glass or reference Repair wiring check for coated lead open or plugged liquid ...

Страница 19: ...nated Section 4 2 1 thru 4 2 2 Defective preamplifier Check preamplifier as instructed in Section 4 4 1 and replace if defective Bias control in analyzer Consult analyzer manual improperly set or mistakenly adjusted Model 1002 1003 only 4 4 MAINTENANCE This section consists of replacing defective parts and making necessary checks to determine if the sensor is operational The following determines i...

Страница 20: ...uF process Cracked glass failure Broken or cracked glass Replace sensor 7MBTT gBJM Input voltage high pH input shorted or sensor Check wiring Replace sensor if Input voltage low miswired necessary Old glass warning 1 Glass electrode worn out 1 Replace sensor 2 Sensor not immersed 2 Be sure electrode measuring tip is in 7MbTT WBso process Reference offset err Reference electrode poisoned Replace se...

Страница 21: ...et the liquid junction on both reference cells and touch them together at the flat surface of the liquid junctions make sure there is direct contact and the surfaces are wet 3 Meter should indicate less than 10 millivolts 4 If meter does not indicate less than 10 millivolts repair the reference cell as instructed in Section 4 2 1 B To check the cell using a laboratory reference electrode pro ceed ...

Страница 22: ...17 MODEL 320HP SECTION 4 0 MAINTENANCE AND TROUBLESHOOTING FIGURE 4 3 Glass Electrode Test Method FIGURE 4 4 Temperature Compensator Check Set Up ...

Страница 23: ...ug must remain porous To check it out remove the plug and top off the reference cell to the brim with deionized water Dry the plug and screw it back into the cell while watching the tip DO NOT use thread sealants other than TEFLON tape on the threads of the plug Water beads should be forced through the foam not just around the edges or threads If it looks clogged replace it 4 4 5 Automatic Tempera...

Страница 24: ...tion Wood Foam KYNAR 1 5 2001553 Glass Electrode w BNC options 50 54 1 5 23642 00 Glass Electrode wo BNC options 55 56 57 1 6 2001779 Housing for pH Electrode 1 23722 00 Pipe mounting assembly optional 1 8 22994 00 Micro junction Reference Element 1 23557 00 Preamplifier for use with Models 54 54e 3081 4081 5081 Xmt 1 22698 02 Preamplifier for use with Models 1181 pH and 1050 1 22698 03 Preamplifi...

Страница 25: ...20 MODEL 320HP SECTION 5 0 PARTS LIST FIGURE 5 1 Model 320HP Front and Side Views Replacement Component Locator DWG NO REV 4320HP06 A ...

Страница 26: ...Carefully package the materials and enclose your Letter of Transmittal and the completed copy of the Return of Materials Request form If possible pack the materials in the same manner as it was received IMPORTANT Please see second section of Return of Materials Request Form Compliance to the OSHA requirements is mandatory for the safety of all personnel MSDS forms and a certification that the inst...

Страница 27: ...___________ ______________________________________________________________________________________________________ ______________________________________________________________________________________________________ WARRANTY REPAIR REQUESTED YES REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO ________________________________________ CUSTOMER PURCHASE ORDER NO ___________________________________...

Страница 28: ...greed upon in writing in advance by an authorized Seller representative All costs of dismantling reinstallation and freight and the time and expenses of Seller s per sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the or...

Страница 29: ...N LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http www raihome com Specifications subject to change without notice Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine CA 92606 USA Tel 949 757 8500 Fax 949 474 7250 http www raihome com Rosemount Analytical Inc 2006 ...

Страница 30: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Отзывы: