![Emerson Rosemount 226 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/emerson/rosemount-226/rosemount-226_instruction-manual_100977005.webp)
1
MODEL 226
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 GENERAL.
The Model 226 Toroidal Conductivity
Sensor greatly reduces the adverse effects of coating
and chemical attack associated with exposed elec-
trodes used in conventional conductivity measurement.
This sensor is available in highly cor ro sion-resistant
PEEK (polyetheretherketone).
PEEK is not recommended for greater than 50% con-
centrations (at 25°C, 77°F) of H
2
SO
4
, HNO
3
, and
H
3
PO
4
. PEEK is not recommended for HF at all.
The Model 226 features rigid construction that will
stand up to high vibration applications such as oil well
drilling and mud shaker trays. A single piece 304SS
tube supports the toroid coils and reinforces the
threaded mounting shaft. This subassembly is then
molded in chemically resistant plastic. There are no
seams or welds that can crack and cause leakage
and subsequent failure.
1.2 MODEL 226 SPECIFICATIONS.
Minimum Range w/1181T :
0 to 80 µS/cm (Code 50)
Maximum Range w/1181T:
0 to 160 mS/cm (Code 50)
Minimum Range w/1054/2054 Series :
0 to 50 µS/cm
Maximum Range w/54C, 1054B, 2081T, 81T, 3081T,
4081T, 1054/1054A/1054B series, and 2054 series:
0 to 2000 mS/cm
Type of Installation:
Submersion/Insertion
Maximum Temperature:
120°C (248°F)
Maximum Pressure:
2035 kPa abs (295 psig)
Materials of Construction:
PEEK (Polyetheretherketone)
Ethylene Propylene Gasket
Integral Cable Length:
20 ft (6.1 m)
Maximum Cable Length with 1181T :
100 ft (30.5 m).
Consult factory for special sensor.
Maximum Cable Length with 1054B/2054 Series:
200 ft (61.0 m)
Process Connections:
1-in. MNPT and 7/8-in. - 9 UNC for
flange mounting
Weight/Shipping Weight: