![Emerson Rosemount 2230 Скачать руководство пользователя страница 109](http://html.mh-extra.com/html/emerson/rosemount-2230/rosemount-2230_reference-manual_103865109.webp)
99
Reference Manual
00809-0100-2230, Rev CA
July 2017
5.6.2
Recommended actions
The FD_RECOMMEN_ACT and RECOMMENDED_ACTION parameters display text strings that will give a recommended course of
action to take based on which type and which specific event of the alert that is active.
provides recommended actions
for Field Diagnostic Alerts as given by the alert default setting for the Rosemount 2230 Display.
Table 5-11. Recommended actions for Field Diagnostics Alerts.
Alert Type
Description
Recommended Action
Failure
Software Failure
1. Restart communication.
2. Load default database to the device and reconfigure
the device.
3. Contact Rosemount Tank Gauging Service
Department.
Database Error
1. Restart communication.
2. Load default database to the device.
3. Reconfigure the device.
Electronics failure - Main Board
Replace the device.
Electronics failure - FF I/O Board
Replace the device.
Memory Failure - FF I/O Board
1. Restart communication.
2. If error persists it may indicate a faulty memory chip.
Replace the device.
Internal Communication Failure
1. Restart communication.
2. Replace the device.
Out of
Specification
Internal Temperature Out of
Limits
Check ambient temperature at installation site.
Invalid Model Code
Contact Emerson Automation Solutions / Rosemount
Tank Gauging service department.
MAO Fault State Mode enabled
One or several Multiple Analog Output (MAO) blocks are
configured with fault-state mode enabled. With this
configuration, the output on the display will show the
fault-state data not the actual data from the connected
devices.
Turn fault state mode off in MAO block.
Function
Check
Check Function
One or more transducer blocks are in Out of Service
mode. Return transducer block to Auto mode.
Содержание Rosemount 2230
Страница 1: ...Reference Manual 00809 0100 2230 Rev CA July 2017 Rosemount 2230 Graphical Field Display ...
Страница 2: ......
Страница 4: ...ii Reference Manual 00809 0100 2230 Rev CA Title Page July 2017 Title Page ...
Страница 16: ...6 Reference Manual 00809 0100 2230 Rev CA Introduction July 2017 Introduction ...
Страница 28: ...18 Reference Manual 00809 0100 2230 Rev CA Overview July 2017 Overview ...
Страница 44: ...34 Reference Manual 00809 0100 2230 Rev CA Installation July 2017 Installation ...
Страница 120: ...110 Reference Manual 00809 0100 2230 Rev CA Service and Troubleshooting July 2017 Service and Troubleshooting ...
Страница 132: ...122 Product Certifications Product Certifications July 2017 Reference Manual 00809 0100 2230 Rev CA ...
Страница 154: ...144 Reference Manual 00809 0100 2230 Rev CA Index July 2017 Index ...
Страница 155: ......