Reference Manual
00809-0100-4828, Rev EA
November 2007
2-3
Rosemount 1595
Liquid or Steam in Horizontal Pipes
The electronics should be mounted below the pipe to ensure that gases do
not collect on the transmitter sensing diaphragms.
Figure 2-2. Liquid and Steam in
Horizontal Pipes
Vertical Pipe Installation
Gas in Vertical Pipes
Mount the electronics above the pipe with the instrument lines sloping down.
Figure 2-3. Gas in Vertical Pipes
FLOW
Flow
Содержание Rosemount 1595
Страница 2: ......
Страница 4: ......
Страница 6: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 TOC 2 ...
Страница 8: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 1 2 ...
Страница 18: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 2 10 ...
Страница 19: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 2 11 ...
Страница 20: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 2 12 ...
Страница 30: ...Reference Manual 00809 0100 4828 Rev EA November 2007 Rosemount 1595 B 2 ...
Страница 31: ......