Reference Manual
00809-0100-4631, Rev BA
April 2007
5-3
Rosemount 1154 Series H
1.
Disengage the header assembly board (4) as described in Step 4 of
the electrical housing disassembly procedure on page 5-3. You need
not remove the module shroud from the electrical housing for
checkout.
2.
Jump connections 1 and 2 on the header assembly board (see Figure
5-1).
3.
Using a low-voltage ohmmeter, check the resistance between the
jumper wire and the module shroud. This resistance should be
greater than 10 M
Ω
. Remove the jumper wire.
4.
Jump connections 3 and 4 on the header assembly board and repeat
Step 3 (see Figure 5-1).
NOTE
The above procedure does not completely test the sensing module. If circuit
board replacement does not correct the abnormal condition, and no other
problems are obvious, the transmitter should be replaced.
DISASSEMBLY
PROCEDURE
NOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
NOTE
The Rosemount 1154 Series H Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
Electrical Housing
Disassembly
1.
The signal terminals and the test terminals are accessible by
unscrewing the cover (1) on the terminal side. This compartment is
identified as “Terminal Side” on the nameplate. The terminals are
permanently attached to the housing and must not be removed.
2.
Circuit boards are located in a separate compartment identified as
“Circuit Side” on the nameplate. Remove power from the transmitter
before removing the circuit cover. Unscrew the cover (1) on the circuit
side to access the circuit boards. A special cover wrench (Rosemount
Part Number 01153-0382-0001) is available from Rosemount to
remove and replace the housing covers.
3.
Unplug the amplifier board (6) after removing 3 holding screws (7).
4.
The header assembly board (4) is permanently attached to the sensor
module within the module shroud (8) and contains the
temperature-compensating resistors. Carefully pull this board off the
bayonet pins and rotate the board 180° about the axis formed by the
connecting leads. This allows access to the calibration board (5).
5.
Disconnect the calibration board (5) by aligning the zero and span
adjust screws so that their slots are perpendicular to the board.
Remove the board by inserting a 6-32 screw in the rivnut on the
board, and carefully pull the board off the bayonet pins.
6.
If replacement of the zero and span adjustment screws (10) is
necessary, remove the nameplate (13) and detach the snap rings (12)
inside the housing.
Содержание Rosemount 1154DH
Страница 6: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Страница 10: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H TOC 4 ...
Страница 12: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Страница 52: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 6 10 ...
Страница 55: ......