Instruction Manual
748467-A
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Installation 2-11
Model MicroCEM
c. Interface Connections
Connection
Designator
Shell
Size
No.
Contacts
AWG
Table
AC Power
J1
12
3
16
Analog Interface
J2
12
22
26, 24, 22
Digital Interface
J3
14
37
26, 24, 22
RS232
J4
10
13
28, 26, 24
LAN Interface
J5
8
6
28, 26, 24, 22
RS485
J6
8
3
24, 22, 20
Phone Line (Modem)
J7
8
3
24, 22, 20
Antenna (Peltier Power)
J8
16
3
14, 12
Table 2-1. Interface Connections
SIGNAL NAME
DEFINITION
PIN
L1
A
L2
85-264 VAC, 47-440 Hz
C
GND
AC Ground
B
Table 2-2. AC Power Connection Terminal Assignments
SIGNAL NAME
DEFINITION
PIN
O
2
CL+
1
O
2
CL-
O
2
Reading, 4-20 mA Output
2
COCL+
3
COCL-
CO Reading, 4-20 mA Output
4
NO
X
+
5
NO
X
-
NO
X
Reading, 4-20 mA Output
6
7
EXP1CL-
External process No. 1, Current Loop input, 4-20 mA
8
9
EXP2CL-
External process No. 2, Current Loop input, 4-20 mA
10
BAROP+
11
BAROP-
Barometric pressure Compensator 0-10V input
12
Table 2-3.
Analog Output Terminal Assignments