Quick start
Topics covered in this chapter:
•
•
•
Make a startup connection to the transmitter
•
Characterize the flowmeter (if required)
•
•
2.1
Power up the meter
The meter must be powered up for all configuration and commissioning tasks, or for
process measurement.
1.
Ensure that all device covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers
and seals are tightly closed. For hazardous area installations, applying power while
housing covers are removed or loose can cause an explosion.
2.
Turn on the electrical power at the power supply.
Postrequisites
Although the sensor is ready to measure shortly after power-up, the electronics can take
up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if
power has been off long enough to allow components to reach ambient temperature,
allow the electronics equalize for approximately 10 minutes.
2.2
Check meter status
Check the meter for any error conditions that require user action or that affect
measurement accuracy.
1.
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions.
2.
Connect to the transmitter and check for active alerts.
Quick start
Configuration and Use Manual
7
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...