Appendix A
Default values and ranges
Topics covered in this appendix:
A.1
Default values and ranges
See Table A-1 for the default values and ranges for the most frequently used transmitter settings.
Table A-1
Transmitter default values and ranges
Type
Setting
Default
Range
Comments
Flow direction
Forward
Flow damping
0.8 sec
0.0 – 40.96
sec
User-entered value is corrected
to nearest lower value in list
of preset values. For gas
applications, Micro Motion
recommends a minimum value
of 2.56.
Mass flow units
g/s
Mass flow cutoff
0.0 g/s
Recommended setting is 5%
of the sensor’s rated maximum
flowrate.
Volume flow type
Liquid
Volume flow units
L/s
Flow
Volume flow cutoff
0/0 L/s
0.0 – x L/s
x is obtained by multiplying the
flow calibration factor by 0.2,
using units of L/s.
Mass factor
1
Density factor
1
Meter factors
Volume factor
1
Configuration and Use Manual
203
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...