Function
Take the wire from terminal
pin:
...and, connect it to the CDM output ter-
minal:
Temperature
Signal converter/flow comput-
er terminals for temperature
(See the signal converter/flow
computer manual for more in-
formation on the device termi-
nals.)
Two wires from the signal converter/flow
computer connect to:
•
mA 1 +
•
mA 1 –
Important
You must configure the mA output 1 to
Temperature to provide this measure-
ment.
Power supply
Signal converter/flow comput-
er terminals for power
(See the signal converter/flow
computer manual for more in-
formation on the device termi-
nals.)
Two wires from the signal converter/flow
computer connect to:
•
PWR +
•
PWR –
4.3.2
Rewire from 7835/7845 Advanced electronics meter
To wire from an existing 7835/7845 meter with Advanced electronics, you must have
purchased one of the following CDM transmitter versions:
•
Analog
•
Time Period Signal (TPS)
•
Any option with Modbus/RS-485
•
Processor for remote-mount 2700 FOUNDATION
™
fieldbus
Procedure
See the following for information on the terminal-to-output relationships from the
7835/7845 to the CDM.
Important
Depending on whether you are wiring in an explosion proof/flameproof or intrinsically safe area, see
the additional information documented for these types of installations.
Function
Take the wire from terminal
pin:
...and, connect it to the CDM output ter-
minal:
Time period or sta-
tus output
(1)
Pin 1
TPS +
Pin 2
TPS –
Analog 1
Pin 3
mA 1 +
Pin 4
mA 1 –
Wiring
Installation Manual
21