![Emerson Micro Motion 5700 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/emerson/micro-motion-5700/micro-motion-5700_installation-manual_104641032.webp)
4.6
Wire the Discrete Output
Use this section to wire the Discrete Output in explosion-proof, nonincendive, or
nonhazardous installations.
Important
Meter installation and wiring should be performed only by suitably-trained personnel.
4.6.1
Wire the Discrete Output (internally powered)
Use this procedure to wire the internally-powered Discrete Output for Channel B or C.
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-16: Discrete Output wiring (internally powered)
A
D
C
B
A. Discrete Output
B. Channel B or C
C. See
for output amplitude versus load resistance
D. Counter
Figure 4-17: Internally powered Discrete Output: output amplitude versus load
resistance [24 VDC (Nom) open circuit]
0
2
4
6
8
10
12
14
16
18
20
22
24
0
750
1500
2250
3000
A
B
A. Output amplitude (V)
B. Load resistor (Ω)
Wiring the channels
Installation Manual
February 2020
MMI-20027478
32