Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
-
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
•
Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
Procedure
See
D3 or D3 and D4 density calibration using ProLink III
Figure 11-11:
Enter density of D3 fluid
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
Done
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 3
Close
Start Calibration
Enter density of D4 fluid
D4 calibration
Fill sensor with D4 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 4
Close
Start Calibration
11.6.3
Perform a D3 or D3 and D4 density calibration using the
Field Communicator
Prerequisites
•
During density calibration, the sensor must be completely filled with the calibration
fluid, and flow through the sensor must be at the lowest rate allowed by your
application. This is usually accomplished by closing the shutoff valve downstream
from the sensor, then filling the sensor with the appropriate fluid.
Measurement support
Configuration and Use Manual
201
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......