SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water. The system
consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the table to verify that you
have received the parts for the option you ordered.
(1)
The analyzer model number is printed on a label attached to the side of the instrument. For example, if you
ordered a panel mount, dual input analyzer, with alarm board, the label should read: T1055-10-22-60.
2.2 INSTALLATION — ANALYZER
2.2.1 General Information
1.
Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme tem-
peratures.
2.
Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are min-
imized or absent.
3.
Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer.
4.
Do not run AC power and relay wiring through the top conduit openings. Keep AC power and relay
wiring separate from other wiring in the analyzer after installation.
5.
The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
6.
See Section 3.1 for removal of conduit knockouts.
7.
To reduce the likelihood of stress on wiring connections, remove the hinged front panel (-11 models) from the
base during wiring installation. Allow sufficient wire length to avoid stress on conductors.
Item
Model/part number
(1)
Analyzer-panel mount
(1)
T1055-[10]-[ ]
Analyzer-pipe/wall mount
T1055-[11]-[ ]
Analyzer-single input
T1055-[ ]-[21]
Analyzer-dual input
T1055-[ ]-[22]
Analyzer- alarm board input
T1055-[ ]-[ ]-[60]
Sensor-EPA standard
8-0108-0002-EPA
Sensor-ISO standard
8-0108-0003-ISO
Cable-3 ft (0.9 m)
24138-00
Cable-20 ft (6.1 m)
24097-00
Cable-50 ft (15.2 m)
24098-00
Calibration cup
24101-00
Molded chamber/debubbler
24170-00
Type of Mounting
Section
Panel
2.2.2
Pipe
2.2.3
Surface
2.2.4
MODEL CLARITY II
SECTION 2.0
INSTALLATION
3