b. Verify
Show join key
has a check mark.
Note
The most common cause of incorrect operation is that the Network ID or Join Key
are not set correctly in the device.
4.4
Device operating limits
Table 4-2: Standard, low-power sensor operational ranges
Sensor
Range
Frequency Response
Vibration Overall X
16g's
2.7 KHz
Vibration Overall Y
16g's
2.7 KHz
Vibration Overall Z
16g's
0.6KHz
PeakVue Z
80g's
20KHz
Temperature
-40
⁰
C to 85
⁰
C
N/A
4.5
Device variable alert limits
This table shows the alert limits of the device variables.
Table 4-3: Overall X
Axis
Severity
Upper Limit
Lower limit
Unit
Remarks
Overall X
Saturated
70
0
in/s
Overall X
Sensor
70
0
in/s
Overall X Out of Range
Overall X
Range
5
0
in/s
Overall X
Fail
1
NaN DV
Unused
in/s
Overall X Out of Range
Overall X
Maint
0.35
NaN DV
Unused
in/s
Overall X Maintenance
Overall X
Advise
0.14
NaN DV
Unused
in/s
Overall X Advise
Table 4-4: Overall Y
Axis
Severity
Upper Limit
Lower Limit
Unit
Remarks
Overall Y
Saturated
70
0
in/s
Overall Y
Sensor
70
0
in/s
Overall Y Out
of Range
Overall Y
Range
5
0
in/s
Overall Y
Fail
1
0
in/s
Overall Y
Failure
User Guide
Operation
MHM-97927-PBF
July 2020
MHM-97927-PBF, DRAFT
47
Содержание AMS
Страница 1: ...User Guide MHM 97927 PBF Rev DRAFT July 2020 AMS Wireless Vibration Monitor User Guide ...
Страница 4: ...Contents User Guide July 2020 MHM 97927 PBF iv MHM 97927 PBF DRAFT ...
Страница 12: ...Introduction User Guide July 2020 MHM 97927 PBF 12 MHM 97927 PBF DRAFT ...
Страница 66: ...Specifications and reference data User Guide July 2020 MHM 97927 PBF 66 MHM 97927 PBF DRAFT ...
Страница 74: ...Product certifications User Guide July 2020 MHM 97927 PBF 74 MHM 97927 PBF DRAFT ...
Страница 77: ...User Guide MHM 97927 PBF July 2020 MHM 97927 PBF DRAFT 77 ...