Micro Motion 7829 Viscomaster Dynamic Viscosity Meter

No Field Calibration Required

The 7829 Viscomaster Dynamic viscosity meter is 
factory-calibrated and no field calibration is 
required. The calibration is traceable to UK National 
Standards through the Micro Motion onsite 
accredited laboratory.

Integral Electronics for Signal Processing

A configured microprocessor-based electronic 
module provides the full signal conditioning, 
calculation, and diagnostic facilities within these 
meters. The 7829 Viscomaster Dynamic does not 
require remote electronics for signal processing. 
This meter simultaneously measures real-time 
viscosity and temperature. The integrally mounted 
electronics can be configured to output a number of 
user-defined parameters.

RS-485/Modbus Communications Interface

All measurements are available digitally through the 
built-in RS-485/Modbus communications interface – 
including calculated density and calculated 
kinematic viscosity at operating temperature – for 
integration into plant data systems. You can use the 
ADView or ProLink II (v2.9 or later) software tools to 
configure or diagnose a meter via the RS-485/
Modbus interface. See “ADView Software Tool 
Features” and “ProLink II Software Tool Features” 
for more information.

User-Configurable 4–20 mA Outputs

Two 4–20 mA outputs are available:

mA output 1 is configurable to output 
parameters such as dynamic or kinematic 
viscosity. 

mA output 2 is set to output temperature, which 
you can configure for zero and span. 

Any of a range of output parameters can be used to 
drive the two 4–20 mA outputs, enabling them to be 
used as the process variable in control applications 
without the need for additional processing 
electronics.

Following are the default settings for the two 
mA outputs.

Code

mA output 1

mA output 2

Viscosity (cSt)

Temp (°C)

4 mA

20 mA

4 mA

20 mA

H

0

25

0

150

J

0

50

0

150

Z

Custom

Features

The Viscomaster Dynamic offers features that are suited to specific applications and industry needs.

Process Requirement

7829 Viscomaster Dynamic

Dynamic viscosity

9

Kinematic viscosity

9

Temperature measurement

9

Integrated density output

Calculated

Sensor type

Fork

Mounting arrangement

Bypass / Inline

Transmitter version

9

mA outputs

Two 4–20 mA

(1 configurable / 1 fixed)

Measurable liquid types

Heavy fuel oil

Principle markets

Marine/Power

Содержание 7829FEANAHRABA

Страница 1: ...cy viscosity measurement Multi variable measurement of dynamic viscosity kinematic viscosity and temperature Unique direct insertion design Integral mount transmitter with analog and digital communications Greatest installation flexibility Continuous real time measurement in various installation configurations Retrofit kits available for easy replacement of existing viscosity meter technologies Su...

Страница 2: ...s Register GL DNV ABS and BV For a complete list of marine approvals see Marine approval classifications on page 7 Advantages Fully integrated fit and forget digital viscosity and temperature measurement for monitoring and control Two 4 20 mA outputs of kinematic viscosity and temperature Modbus RS 485 communications output of all parameters including calculated density and calculated kinematic vi...

Страница 3: ...asured by detecting the bandwidth around the fork s natural vibration frequency The following figure illustrates how the 7829 meter uses the frequencies at point A the lower 3 db point and at point B the upper 3 db point to digitally calculate bandwidth and quality factor Quality factor is then used to define fluid viscosity using standard calibration techniques Product 1 Product 2 Response amplit...

Страница 4: ...II v2 9 or later software tools to configure or diagnose a meter via the RS 485 Modbus interface See ADView Software Tool Features and ProLink II Software Tool Features for more information User Configurable 4 20 mA Outputs Two 4 20 mA outputs are available mA output 1 is configurable to output parameters such as dynamic or kinematic viscosity mA output 2 is set to output temperature which you can...

Страница 5: ...communicates with the meter through a standard serial port and provides the following functionality Setting up a serial link to communicate with the 7829 transmitter Configuring the 7829 transmitter Displaying data real time or as a graph Logging data to a file Verifying correct operation of the system and diagnosing faults ProLink II Software Tool Features The ProLink II software tool v2 9 or lat...

Страница 6: ...lity 0 5 of reading Process Ambient 40 F to 392 F 40 C to 200 C 40 F to 185 F 40 C to 85 C Technology 100 Ohms RTD 4 wire Accuracy PT100 BS1904 Class B DIN 43760 Class B Maximum operating pressure 1 1 Actual maximum operating pressures are limited by the process connection rating 3000 psi 207 bar Test pressure Tested to 1 5 x the maximum operating pressure PED compliance Outside the scope of Europ...

Страница 7: ...nvironment Weather rating IP66 ATEX ATEX approved Certification for use in Europe ATEX II2G Ex d IIC T4 CSA C US CSA approved Certification for use in Canada and USA Class I Division 1 Groups C D T4 Marine Approval Country American Bureau of Shipping USA Lloyds London United Kingdom Germanische Lloyd Germany Det Norske Veritas Norway Bureau Veritas France RINA Italy Nippo Kaiji Kyokai Japan Russia...

Страница 8: ...aint finish Weight 7 5 lb 3 4 kg typical Power supply requirement For transmitter For mA outputs 20 to 28 VDC 50 mA 22 mA per output Outputs mA Accuracy Repeatability Out of range System alarm Two externally powered 4 20 mA 0 1 of reading 0 05 of full scale at 68 F 20 C 0 05 of full scale over range 40 F to 185 F 40 C to 85 C 3 9 or 20 8 mA on 4 20 mA 2 or 22 mA on 4 20 mA Programmable alarm state...

Страница 9: ... Motion 7829 Viscomaster Dynamic Viscosity Meter 9 Dimensions Dimensions in inches mm 4 9 125 4 5 5 140 4 3 109 Mounting to suit 1 5 Swagelok or similar cone 1 2 31 1 2 NPT both sides 4 8 121 0 5 12 1 38 35 ...

Страница 10: ...sist of a number of supply pumps fed by either HFO or MDO a flow meter in line filters to remove the impurities and a holding mixing tank Following the supply section the fuel is usually sent to booster pumps that increase the flow rate up to a maximum of 20 m3 hr and then through a series of heat exchangers liquid or electric to change the product viscosity for efficient combustion Following is a...

Страница 11: ...no stem extension and with standard spigot Code Default software configuration of 4 20 mA output 1 1 1 4 20 mA output 2 default setting is temperature H 0 to 25 cSt J 0 to 50 cSt Z ETO custom configuration Code Calibration range R 0 5 to 50 cP no density calibration Code Calibration type A Free stream B 2 schedule 40 boundary 200 cP limit E 3 schedule 80 boundary 2 2 Compatible with DN80 XS pipewo...

Страница 12: ...by more than 750 000 meters installed worldwide and over 30 years of flow and density measurement experience 2013 Micro Motion Inc All rights reserved The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co Micro Motion ELITE MVD ProLink MVD Direct Connect and PlantWeb are marks of one of the Emerson Process Management family of companies All other trademarks are...

Отзывы: