8
Micro Motion 7829 Viscomaster Dynamic Viscosity Meter
Materials of construction
Weight
Electrical
Wetted parts
316L Stainless steel
Tine finish
PFA laminated
(1)
(1) PFA is applied only to the tines for its anti-stick properties not for corrosion protection.
Electronics enclosure
Sand cast low copper alloy
Polyurethane paint finish
Weight
7.5 lb (3.4 kg) typical
Power supply requirement
For transmitter
For mA outputs
20 to 28 VDC, 50 mA
22 mA per output
Outputs
mA
Accuracy
Repeatability
Out-of-range
System alarm
Two externally powered 4–20 mA
±0.1% of reading, ±0.05% of full scale at 68 °F (20 °C)
±0.05% of full scale, over range –40 °F to +185 °F
(–40 °C to +85 °C)
3.9 or 20.8 mA on 4–20 mA
2 or 22 mA on 4–20 mA (Programmable alarm state)
Communications
RS-485 (Modbus)
Electrical connection
Screw terminal, cable entry to suit ½" NPT gland (20 mm adaptor available)